SINO calculator

Absolute SINOs

The table below displays the absolute signal-to-noise (SINO) ratios for several standard nuclei. The SINO values correspond to the latest values recorded, not the original factory specifications.

Magritek Spinsolve 80 Carbon Bruker Avance 300 Bruker Avance III HD 400 Bruker Avance II+ 600 Bruker Avance Neo 600
HCF BBO5 BBFO5 BBO5 TBO5 TCI5
1H 13 94 590 400 843 3730
13C 6 104 219 357 397 613
15N N/A 21 5 18 59 29
19F 3 N/A 288 N/A 1444 4000
29Si N/A 47 99 153 177 N/A
31P N/A 131 141 128 406 N/A

1 Acquisition parameters from Magritek were translated and recalculated into their corresponding Bruker counterparts.
2 Standard reference sample purchased from Bruker Corporation.
Magritek Spinsolve 80 Carbon1 Bruker Avance 300 Bruker Avance III HD 400 Bruker Avance II+ 600 Bruker Avance Neo 600
HCF BBO5 BBFO5 BBO5 TBO5 TCI5
Sample2 0.1 % ethylbenzene in CDCl3
PULPROG zg
TD 32768 65536 32768
NS 1
DS 0
SWH (Hz) 2997.602 4000.000 6009.615 5882.353
AQ (sec) 5.4657526 8.1920004 2.7263477 2.7852800
DW (µsec) 166.800 125.000 83.200 85.000
DE (µsec) 6.00 6.50 10.00
TE (K) 298.0
D1 (sec) 60.00000000 113.57360077
SFO1 (MHz) 300.1312005 400.1716007 600.1324005 600.3524014
NUC1 1H
SI 16384 131072 16384
WDW EM
LB (Hz) 1.00
PC 1.00
Signal range (ppm) 3 2
Noise range (ppm) 6 3
Noise width (Hz) 200
1 Acquisition parameters from Magritek were translated and recalculated into their corresponding Bruker counterparts.
2 Standard reference sample purchased from Bruker Corporation.
Magritek Spinsolve 80 Carbon1 Bruker Avance 300 Bruker Avance III HD 400 Bruker Avance II+ 600 Bruker Avance Neo 600
HCF BBO5 BBFO5 BBO5 TBO5 TCI5
Sample2 40 % 1,4-dioxane in C6D6 (ASTM)
PULPROG zg
TD 131072 65536
NS 1
DS 0
SWH (Hz) 50000.000 15060.241 20161.291 29761.904 30120.482
AQ (sec) 1.3107200 2.1758451 1.6252928 1.1010548 1.0878977
DW (µsec) 20.000 33.200 24.800 16.800 16.600
DE (µsec) 6.00 6.50 6.89 19.34
TE (K) 299.65 298.0
D1 (sec) 300.00000000 828.91210938
SFO1 (MHz) 20.2384611 75.4677190 100.6328888 150.9028090 150.9732215
NUC1 13C
SI 131072 32768 65536 32768 131072
WDW EM
LB (Hz) 3.50
PC N/A 1.40
Signal range (ppm) 140 120
Noise range (ppm) 120 70
Noise width (ppm) 40
1 Acquisition parameters from Magritek were translated and recalculated into their corresponding Bruker counterparts.
2 Standard reference sample purchased from Bruker Corporation.
Magritek Spinsolve 80 Carbon1 Bruker Avance 300 Bruker Avance III HD 400 Bruker Avance II+ 600 Bruker Avance Neo 600
HCF BBO5 BBFO5 BBO5 TBO5 TCI5
Sample2 N/A 90 % formamide in DMSO-d6
PULPROG N/A zgig
TD N/A 32768
NS N/A 1
DS N/A 0
SWH (Hz) N/A 609.162 811.162 1216.545 1190.476
AQ (sec) N/A 26.8960247 20.1981945 13.4676981 13.7625599
DW (µsec) N/A 820.800 616.400 411.000 420.000
DE (µsec) N/A 6.50 18.00
TE (K) N/A 298.0
D1 (sec) N/A 120.00000000 10.00000000 116.23739624
D11 (sec) N/A 0.03000000
SFO1 (MHz) N/A 30.4153393 40.5534357 60.8174862 60.8397761 60.8397761
NUC1 N/A 15N
SFO2 (MHz) N/A 300.1321909 400.1729212 600.1343810 600.3543826 600.3543826
NUC2 N/A 1H
CPDPRG2 N/A waltz16 waltz64
SI N/A 16384
WDW N/A EM
LB (Hz) N/A 0.30
PC N/A 1.00
Signal range (ppm) N/A 113-111
Noise range (ppm) N/A 120-113
Noise width (ppm) N/A 2
1 Acquisition parameters from Magritek were translated and recalculated into their corresponding Bruker counterparts.
2 19F tuned on BB coil and recorded via BB channel.
3 19F can also be recorded via separate 19F channel.
4 19F tuned on 1H coil and recorded via 1H channel.
5 Standard reference sample purchased from Bruker Corporation.
6 Pulse-acquire sequence during which the external 19F lock sample is saturated before the reference sample is excited to reduce lock interference.
7 To due insufficient sensitivity for a single scan experiment, 16 scans were recorded and the corresponding SINO for a single scan was calculated from this value.
Magritek Spinsolve 80 Carbon1 Bruker Avance 300 Bruker Avance III HD 400 Bruker Avance II+ 600 Bruker Avance Neo 600
HCF BBO5 BBFO52 BBO5 TBO52, 3 TCI54
Sample5 0.05 % α,α,α-trifluorotoluene in CDCl3 N/A 0.05 % α,α,α-trifluorotoluene in CDCl3 N/A 0.05 % α,α,α-trifluorotoluene in CDCl3
PULPROG zg6 N/A zg N/A zg
TD 65536 N/A 32768 N/A 32768
NS 17 N/A 1 N/A 1
DS 0 N/A 0 N/A 0
SWH (Hz) 10000.000 N/A 3765.060 N/A 5555.556
AQ (sec) 3.2768000 N/A 4.3515906 N/A 2.9491200
DW (µsec) 100.000 N/A 132.800 N/A 90.000
DE (µsec) N/A 6.50 N/A 6.50 18.00
TE (K) 299.65 N/A 298.0 N/A 298.0
D1 (sec) 5.00000000 N/A 5.00000000 N/A 32.05088043
SFO1 (MHz) 75.7198312 N/A 376.5122822 N/A 564.8579387
NUC1 19F N/A 19F N/A 19F
SI 65536 N/A 16384 N/A 32768
WDW EM N/A EM N/A EM
LB (Hz) 2.00 N/A 2.00 N/A 2.00
PC N/A N/A 1.00 N/A 1.00
Signal range (ppm) -61 -65 N/A -61 -65 N/A -61 -65
Noise range (ppm) -59 -62 N/A -59 -62 N/A -59 -62
Noise width (ppm) 1 N/A 1 N/A 1
1 Acquisition parameters from Magritek were translated and recalculated into their corresponding Bruker counterparts.
2 Standard reference sample purchased from Bruker Corporation.
Magritek Spinsolve 80 Carbon1 Bruker Avance 300 Bruker Avance III HD 400 Bruker Avance II+ 600 Bruker Avance Neo 600
HCF BBO5 BBFO5 BBO5 TBO5 TCI5
Sample2 N/A 85 % hexamethyldisiloxane in C6D6 N/A
PULPROG N/A zg N/A
TD N/A 32768 N/A
NS N/A 1 N/A
DS N/A 0 N/A
SWH (Hz) N/A 3591.954 4761.905 7142.857 N/A
AQ (sec) N/A 4.5614948 3.4406400 2.2937601 N/A
DW (µsec) N/A 139.200 105.000 70.000 N/A
DE (µsec) N/A 6.50 N/A
TE (K) N/A 298.0 N/A
D1 (sec) N/A 400 416.60620117 N/A
SFO1 (MHz) N/A 59.6277763 79.5030397 119.2297258 119.2734338 N/A
NUC1 N/A 29Si N/A
SI N/A 65536 N/A
WDW N/A EM N/A
LB (Hz) N/A 5.00 N/A
PC N/A 1.40 N/A
Signal range (ppm) N/A 10 5 N/A
Noise range (ppm) N/A 30 10 N/A
Noise width (ppm) N/A 10 N/A
1 Acquisition parameters from Magritek were translated and recalculated into their corresponding Bruker counterparts.
2 Standard reference sample purchased from Bruker Corporation.
Magritek Spinsolve 80 Carbon1 Bruker Avance 300 Bruker Avance III HD 400 Bruker Avance II+ 600 Bruker Avance Neo 600
HCF BBO5 BBFO5 BBO5 TBO5 TCI5
Sample2 N/A 0.0485 M triphenylphosphate in acetone-d6 N/A
PULPROG N/A zg N/A
TD N/A 32768 N/A
NS N/A 1 N/A
DS N/A 0 N/A
SWH (Hz) N/A 6082.725 8081.896 12175.324 11904.762 N/A
AQ (sec) N/A 2.6935797 2.0272470 1.3457226 1.3762560 N/A
DW (µsec) N/A 82.200 61.867 41.067 42.000 N/A
DE (µsec) N/A 6.00 6.50 N/A
TE (K) N/A 298.0 N/A
D1 (sec) N/A 2.00000000 119.62370300 N/A
SFO1 (MHz) N/A 121.4927571 161.9891934 242.9331900 243.0218553 N/A
NUC1 N/A 31P N/A
SI N/A 16384 N/A
WDW N/A EM N/A
LB (Hz) N/A 5.00 N/A
PC N/A 1.40 N/A
Signal range (ppm) N/A -14 -18 N/A
Noise range (ppm) N/A 5 -15 N/A
Noise width (ppm) N/A 5 N/A

Relative SINOs

Performance factors

The table below shows the relative performance factors \(P\) for several standard nuclei. These values can be used to calculate the difference in SINO when changing from one spectrometer to another. To determine the conversion factor, first select the proper conversion table by clicking on the desired nucleus. When changing the acquisition from spectrometer 1 to spectrometer 2, search for the number in the row (horizontal) of spectrometer 1 and the column (vertical) of spectrometer 2. Values higher than 1 mean that for a constant SINO, less scans or lower sample concentrations are required going from spectrometer 1 to spectrometer 2. The opposite holds for performance factors smaller than 1.

Relative performance factors \(P\) for the spectrometers in automation for direct 1H observation. To determine a relative performance factor going from one spectrometer to another, read the former from the vertical header and the latter from the horizontal header. For example, the relative 1H performance factor \(P\) for switching from the BBO5 probe on the Bruker Avance 300 to the TCI5 cryoprobe on the Bruker Avance Neo 600 equals 39.681.
Magritek Spinsolve 80 Carbon Bruker Avance 300 Bruker Avance III HD 400 Bruker Avance II+ 600 Bruker Avance Neo 600
HCF BBO5 BBFO5 BBO5 TBO5 TCI5
Magritek Spinsolve 80 Carbon HCF 1 7.231 45.385 30.769 64.846 286.923
Bruker Avance 300 BBO5 0.138 1 6.277 4.255 8.968 39.681
Bruker Avance III HD 400 BBFO5 0.022 0.159 1 0.678 1.429 6.322
Bruker Avance II+ 600 BBO5 0.033 0.235 1.475 1 2.108 9.325
Bruker Avance Neo 600 TBO5 0.015 0.112 0.700 0.474 1 4.425
TCI5 0.003 0.025 0.158 0.107 0.226 1
Relative performance factors \(P\) for the spectrometers in automation for direct 13C observation. To determine a relative performance factor going from one spectrometer to another, read the former from the vertical header and the latter from the horizontal header. For example, the relative 13C performance factor \(P\) for switching from the BBO5 probe on the Bruker Avance 300 to the BBO5 probe on the Bruker Avance II+ 600 equals 3.433.
Magritek Spinsolve 80 Carbon Bruker Avance 300 Bruker Avance III HD 400 Bruker Avance II+ 600 Bruker Avance Neo 600
HCF BBO5 BBFO5 BBO5 TBO5 TCI5
Magritek Spinsolve 80 Carbon HCF 1 17.333 36.500 59.500 66.167 102.167
Bruker Avance 300 BBO5 0.058 1 2.106 3.433 3.817 5.894
Bruker Avance III HD 400 BBFO5 0.027 0.475 1 1.630 1.813 2.799
Bruker Avance II+ 600 BBO5 0.017 0.291 0.613 1 1.112 1.717
Bruker Avance Neo 600 TBO5 0.015 0.262 0.552 0.899 1 1.544
TCI5 0.010 0.170 0.357 0.582 0.648 1
Relative performance factors \(P\) for the spectrometers in automation for direct 15N observation. To determine a relative performance factor going from one spectrometer to another, read the former from the vertical header and the latter from the horizontal header. For example, the relative 15N performance factor \(P\) for switching from the BBFO5 probe on the Bruker Avance III HD 400 to the BBO5 probe on the Bruker Avance II+ 600 equals 3.600.
Magritek Spinsolve 80 Carbon Bruker Avance 300 Bruker Avance III HD 400 Bruker Avance II+ 600 Bruker Avance Neo 600
HCF BBO5 BBFO5 BBO5 TBO5 TCI5
Magritek Spinsolve 80 Carbon HCF N/A N/A N/A N/A N/A N/A
Bruker Avance 300 BBO5 N/A 1 0.238 0.857 2.810 1.381
Bruker Avance III HD 400 BBFO5 N/A 4.200 1 3.600 11.800 5.800
Bruker Avance II+ 600 BBO5 N/A 1.167 0.278 1 3.278 1.611
Bruker Avance Neo 600 TBO5 N/A 0.356 0.085 0.305 1 0.492
TCI5 N/A 0.724 0.172 0.621 2.034 1
Relative performance factors \(P\) for the spectrometers in automation for direct 19F observation. To determine a relative performance factor going from one spectrometer to another, read the former from the vertical header and the latter from the horizontal header. For example, the relative 19F performance factor \(P\) for switching from the BBFO5 probe on the Bruker Avance III HD 400 to the TBO5 probe on the Bruker Avance Neo 600 equals 5.014.
Magritek Spinsolve 80 Carbon Bruker Avance 300 Bruker Avance III HD 400 Bruker Avance II+ 600 Bruker Avance Neo 600
HCF BBO5 BBFO5 BBO5 TBO5 TCI5
Magritek Spinsolve 80 Carbon HCF 1 N/A 96 N/A 481.333 1333.333
Bruker Avance 300 BBO5 N/A N/A N/A N/A N/A N/A
Bruker Avance III HD 400 BBFO5 0.010 N/A 1 N/A 5.014 13.889
Bruker Avance II+ 600 BBO5 N/A N/A N/A N/A N/A N/A
Bruker Avance Neo 600 TBO5 0.002 N/A 0.199 N/A 1 2.770
TCI5 0.001 N/A 0.072 N/A 0.361 1
Relative performance factors \(P\) for the spectrometers in automation for direct 29Si observation. To determine a relative performance factor going from one spectrometer to another, read the former from the vertical header and the latter from the horizontal header. For example, the relative 29Si performance factor \(P\) for switching from the BBFO5 probe on the Bruker Avance III HD 400 to the BBO5 probe on the Bruker Avance II+ 600 equals 1.545.
Magritek Spinsolve 80 Carbon Bruker Avance 300 Bruker Avance III HD 400 Bruker Avance II+ 600 Bruker Avance Neo 600
HCF BBO5 BBFO5 BBO5 TBO5 TCI5
Magritek Spinsolve 80 Carbon HCF N/A N/A N/A N/A N/A N/A
Bruker Avance 300 BBO5 N/A 1 2.106 3.255 3.766 N/A
Bruker Avance III HD 400 BBFO5 N/A 0.475 1 1.545 1.788 N/A
Bruker Avance II+ 600 BBO5 N/A 0.307 0.647 1 1.157 N/A
Bruker Avance Neo 600 TBO5 N/A 0.266 0.559 0.864 1 N/A
TCI5 N/A N/A N/A N/A N/A N/A
Relative performance factors \(P\) for the spectrometers in automation for direct 29Si observation. To determine a relative performance factor going from one spectrometer to another, read the former from the vertical header and the latter from the horizontal header. For example, the relative 31P performance factor \(P\) for switching from the BBO5 probe on the Bruker Avance II+ 600 to the TBO5 probe on the Bruker Avance Neo 600 equals 3.172.
Magritek Spinsolve 80 Carbon Bruker Avance 300 Bruker Avance III HD 400 Bruker Avance II+ 600 Bruker Avance Neo 600
HCF BBO5 BBFO5 BBO5 TBO5 TCI5
Magritek Spinsolve 80 Carbon HCF N/A N/A N/A N/A N/A N/A
Bruker Avance 300 BBO5 N/A 1 1.076 0.977 3.099 N/A
Bruker Avance III HD 400 BBFO5 N/A 0.929 1 0.908 2.879 N/A
Bruker Avance II+ 600 BBO5 N/A 1.023 1.102 1 3.172 N/A
Bruker Avance Neo 600 TBO5 N/A 0.323 0.347 0.315 1 N/A
TCI5 N/A N/A N/A N/A N/A N/A

For example, if you have performed a 1H experiment on the Bruker Avance 300 with a SINO of 250 and you would perform the same experiment on the Bruker Avance II+ 600, the SINO would be 250 * 4.255 = 1063.75.

Convert SINOs for different spectrometers

Use the applet below to calculate the changes in sample concentration or number of scans when changing from one spectrometer to another. Find the performance factor for the desired conversion in the table below (note the proper nucleus), by clicking on that value in the table. Next, fill in all reference values \(NS_{ref}\), \(SINO_{ref}\) and \(C_{ref}\) and two out of three of the calculated values \(NS_{calc}\), \(SINO_{calc}\) and \(C_{calc}\). The third, blank parameter will then be calculated according to the formula below. If a parameter remains constant, you can fillin the same value for the reference and calculated field, or toggle the corresponding 'Constant' button.

\[\frac{C_{calc}^2 \times NS_{calc}}{SINO_{calc}^2} = P^2 \times \frac{C_{ref}^2 \times NS_{ref}}{SINO_{ref}^2} \]
Please fill in \(NS_{ref}\).
Please fill in \(SINO_{ref}\).
Please fill in \(C_{ref}\).
Please fill in the \(P\) factor.