Absolute SINOs
The table below displays the absolute signal-to-noise (SINO) ratios for several standard nuclei. The SINO values correspond to the latest values recorded, not the original factory specifications.
| 1H |
13 |
94 |
590 |
400 |
843 |
3730 |
| 13C |
6 |
104 |
219 |
357 |
397 |
613 |
| 15N |
N/A |
21 |
5 |
18 |
59 |
29 |
| 19F |
3 |
N/A |
288 |
N/A |
1444 |
4000 |
| 29Si |
N/A |
47 |
99 |
153 |
177 |
N/A |
| 31P |
N/A |
131 |
141 |
128 |
406 |
N/A |
1 Acquisition parameters from Magritek were translated and recalculated into their corresponding Bruker counterparts.
2 Standard reference sample purchased from Bruker Corporation.
| Sample2 |
0.1 % ethylbenzene in CDCl3 |
| PULPROG |
zg |
| TD |
|
32768 |
65536 |
32768 |
| NS |
1 |
| DS |
0 |
| SWH (Hz) |
|
2997.602 |
4000.000 |
6009.615 |
5882.353 |
| AQ (sec) |
|
5.4657526 |
8.1920004 |
2.7263477 |
2.7852800 |
| DW (µsec) |
|
166.800 |
125.000 |
83.200 |
85.000 |
| DE (µsec) |
|
6.00 |
6.50 |
10.00 |
| TE (K) |
298.0 |
| D1 (sec) |
|
60.00000000 |
113.57360077 |
| SFO1 (MHz) |
|
300.1312005 |
400.1716007 |
600.1324005 |
600.3524014 |
| NUC1 |
1H |
| SI |
|
16384 |
131072 |
16384 |
| WDW |
EM |
| LB (Hz) |
1.00 |
| PC |
1.00 |
| Signal range (ppm) |
3 2 |
| Noise range (ppm) |
6 3 |
| Noise width (Hz) |
200 |
1 Acquisition parameters from Magritek were translated and recalculated into their corresponding Bruker counterparts.
2 Standard reference sample purchased from Bruker Corporation.
| Sample2 |
40 % 1,4-dioxane in C6D6 (ASTM) |
| PULPROG |
zg |
| TD |
131072 |
65536 |
| NS |
1 |
| DS |
0 |
| SWH (Hz) |
50000.000 |
15060.241 |
20161.291 |
29761.904 |
30120.482 |
| AQ (sec) |
1.3107200 |
2.1758451 |
1.6252928 |
1.1010548 |
1.0878977 |
| DW (µsec) |
20.000 |
33.200 |
24.800 |
16.800 |
16.600 |
| DE (µsec) |
|
6.00 |
6.50 |
6.89 |
19.34 |
| TE (K) |
299.65 |
298.0 |
| D1 (sec) |
300.00000000 |
828.91210938 |
| SFO1 (MHz) |
20.2384611 |
75.4677190 |
100.6328888 |
150.9028090 |
150.9732215 |
| NUC1 |
13C |
| SI |
131072 |
32768 |
65536 |
32768 |
131072 |
| WDW |
EM |
| LB (Hz) |
3.50 |
| PC |
N/A |
1.40 |
| Signal range (ppm) |
140 120 |
| Noise range (ppm) |
120 70 |
| Noise width (ppm) |
40 |
1 Acquisition parameters from Magritek were translated and recalculated into their corresponding Bruker counterparts.
2 Standard reference sample purchased from Bruker Corporation.
| Sample2 |
N/A |
90 % formamide in DMSO-d6 |
| PULPROG |
N/A |
zgig |
| TD |
N/A |
32768 |
| NS |
N/A |
1 |
| DS |
N/A |
0 |
| SWH (Hz) |
N/A |
609.162 |
811.162 |
1216.545 |
1190.476 |
| AQ (sec) |
N/A |
26.8960247 |
20.1981945 |
13.4676981 |
13.7625599 |
| DW (µsec) |
N/A |
820.800 |
616.400 |
411.000 |
420.000 |
| DE (µsec) |
N/A |
6.50 |
18.00 |
| TE (K) |
N/A |
298.0 |
| D1 (sec) |
N/A |
120.00000000 |
10.00000000 |
116.23739624 |
| D11 (sec) |
N/A |
0.03000000 |
| SFO1 (MHz) |
N/A |
30.4153393 |
40.5534357 |
60.8174862 |
60.8397761 |
60.8397761 |
| NUC1 |
N/A |
15N |
| SFO2 (MHz) |
N/A |
300.1321909 |
400.1729212 |
600.1343810 |
600.3543826 |
600.3543826 |
| NUC2 |
N/A |
1H |
| CPDPRG2 |
N/A |
waltz16 |
waltz64 |
| SI |
N/A |
16384 |
| WDW |
N/A |
EM |
| LB (Hz) |
N/A |
0.30 |
| PC |
N/A |
1.00 |
| Signal range (ppm) |
N/A |
113-111 |
| Noise range (ppm) |
N/A |
120-113 |
| Noise width (ppm) |
N/A |
2 |
1 Acquisition parameters from Magritek were translated and recalculated into their corresponding Bruker counterparts.
2 19F tuned on BB coil and recorded via BB channel.
3 19F can also be recorded via separate 19F channel.
4 19F tuned on 1H coil and recorded via 1H channel.
5 Standard reference sample purchased from Bruker Corporation.
6 Pulse-acquire sequence during which the external 19F lock sample is saturated before the reference sample is excited to reduce lock interference.
7 To due insufficient sensitivity for a single scan experiment, 16 scans were recorded and the corresponding SINO for a single scan was calculated from this value.
| Sample5 |
0.05 % α,α,α-trifluorotoluene in CDCl3 |
N/A |
0.05 % α,α,α-trifluorotoluene in CDCl3 |
N/A |
0.05 % α,α,α-trifluorotoluene in CDCl3 |
| PULPROG |
zg6 |
N/A |
zg |
N/A |
zg |
| TD |
65536 |
N/A |
32768 |
N/A |
32768 |
| NS |
17 |
N/A |
1 |
N/A |
1 |
| DS |
0 |
N/A |
0 |
N/A |
0 |
| SWH (Hz) |
10000.000 |
N/A |
3765.060 |
N/A |
5555.556 |
| AQ (sec) |
3.2768000 |
N/A |
4.3515906 |
N/A |
2.9491200 |
| DW (µsec) |
100.000 |
N/A |
132.800 |
N/A |
90.000 |
| DE (µsec) |
|
N/A |
6.50 |
N/A |
6.50 |
18.00 |
| TE (K) |
299.65 |
N/A |
298.0 |
N/A |
298.0 |
| D1 (sec) |
5.00000000 |
N/A |
5.00000000 |
N/A |
32.05088043 |
| SFO1 (MHz) |
75.7198312 |
N/A |
376.5122822 |
N/A |
564.8579387 |
| NUC1 |
19F |
N/A |
19F |
N/A |
19F |
| SI |
65536 |
N/A |
16384 |
N/A |
32768 |
| WDW |
EM |
N/A |
EM |
N/A |
EM |
| LB (Hz) |
2.00 |
N/A |
2.00 |
N/A |
2.00 |
| PC |
N/A |
N/A |
1.00 |
N/A |
1.00 |
| Signal range (ppm) |
-61 -65 |
N/A |
-61 -65 |
N/A |
-61 -65 |
| Noise range (ppm) |
-59 -62 |
N/A |
-59 -62 |
N/A |
-59 -62 |
| Noise width (ppm) |
1 |
N/A |
1 |
N/A |
1 |
1 Acquisition parameters from Magritek were translated and recalculated into their corresponding Bruker counterparts.
2 Standard reference sample purchased from Bruker Corporation.
| Sample2 |
N/A |
85 % hexamethyldisiloxane in C6D6 |
N/A |
| PULPROG |
N/A |
zg |
N/A |
| TD |
N/A |
32768 |
N/A |
| NS |
N/A |
1 |
N/A |
| DS |
N/A |
0 |
N/A |
| SWH (Hz) |
N/A |
3591.954 |
4761.905 |
7142.857 |
N/A |
| AQ (sec) |
N/A |
4.5614948 |
3.4406400 |
2.2937601 |
N/A |
| DW (µsec) |
N/A |
139.200 |
105.000 |
70.000 |
N/A |
| DE (µsec) |
N/A |
6.50 |
N/A |
| TE (K) |
N/A |
298.0 |
N/A |
| D1 (sec) |
N/A |
400 |
416.60620117 |
N/A |
| SFO1 (MHz) |
N/A |
59.6277763 |
79.5030397 |
119.2297258 |
119.2734338 |
N/A |
| NUC1 |
N/A |
29Si |
N/A |
| SI |
N/A |
65536 |
N/A |
| WDW |
N/A |
EM |
N/A |
| LB (Hz) |
N/A |
5.00 |
N/A |
| PC |
N/A |
1.40 |
N/A |
| Signal range (ppm) |
N/A |
10 5 |
N/A |
| Noise range (ppm) |
N/A |
30 10 |
N/A |
| Noise width (ppm) |
N/A |
10 |
N/A |
1 Acquisition parameters from Magritek were translated and recalculated into their corresponding Bruker counterparts.
2 Standard reference sample purchased from Bruker Corporation.
| Sample2 |
N/A |
0.0485 M triphenylphosphate in acetone-d6 |
N/A |
| PULPROG |
N/A |
zg |
N/A |
| TD |
N/A |
32768 |
N/A |
| NS |
N/A |
1 |
N/A |
| DS |
N/A |
0 |
N/A |
| SWH (Hz) |
N/A |
6082.725 |
8081.896 |
12175.324 |
11904.762 |
N/A |
| AQ (sec) |
N/A |
2.6935797 |
2.0272470 |
1.3457226 |
1.3762560 |
N/A |
| DW (µsec) |
N/A |
82.200 |
61.867 |
41.067 |
42.000 |
N/A |
| DE (µsec) |
N/A |
6.00 |
6.50 |
N/A |
| TE (K) |
N/A |
298.0 |
N/A |
| D1 (sec) |
N/A |
2.00000000 |
119.62370300 |
N/A |
| SFO1 (MHz) |
N/A |
121.4927571 |
161.9891934 |
242.9331900 |
243.0218553 |
N/A |
| NUC1 |
N/A |
31P |
N/A |
| SI |
N/A |
16384 |
N/A |
| WDW |
N/A |
EM |
N/A |
| LB (Hz) |
N/A |
5.00 |
N/A |
| PC |
N/A |
1.40 |
N/A |
| Signal range (ppm) |
N/A |
-14 -18 |
N/A |
| Noise range (ppm) |
N/A |
5 -15 |
N/A |
| Noise width (ppm) |
N/A |
5 |
N/A |
Relative SINOs
Performance factors
The table below shows the relative performance factors \(P\) for several standard nuclei. These values can be used to calculate the difference in SINO when changing from one spectrometer to another. To determine the conversion factor, first select the proper conversion table by clicking on the desired nucleus. When changing the acquisition from spectrometer 1 to spectrometer 2, search for the number in the row (horizontal) of spectrometer 1 and the column (vertical) of spectrometer 2. Values higher than 1 mean that for a constant SINO, less scans or lower sample concentrations are required going from spectrometer 1 to spectrometer 2. The opposite holds for performance factors smaller than 1.
Relative performance factors \(P\) for the spectrometers in automation for direct 1H observation. To determine a relative performance factor going from one spectrometer to another, read the former from the vertical header and the latter from the horizontal header. For example, the relative 1H performance factor \(P\) for switching from the BBO5 probe on the Bruker Avance 300 to the TCI5 cryoprobe on the Bruker Avance Neo 600 equals 39.681.
| Magritek Spinsolve 80 Carbon |
HCF |
1 |
7.231 |
45.385 |
30.769 |
64.846 |
286.923 |
| Bruker Avance 300 |
BBO5 |
0.138 |
1 |
6.277 |
4.255 |
8.968 |
39.681 |
| Bruker Avance III HD 400 |
BBFO5 |
0.022 |
0.159 |
1 |
0.678 |
1.429 |
6.322 |
| Bruker Avance II+ 600 |
BBO5 |
0.033 |
0.235 |
1.475 |
1 |
2.108 |
9.325 |
| Bruker Avance Neo 600 |
TBO5 |
0.015 |
0.112 |
0.700 |
0.474 |
1 |
4.425 |
| TCI5 |
0.003 |
0.025 |
0.158 |
0.107 |
0.226 |
1 |
Relative performance factors \(P\) for the spectrometers in automation for direct 13C observation. To determine a relative performance factor going from one spectrometer to another, read the former from the vertical header and the latter from the horizontal header. For example, the relative 13C performance factor \(P\) for switching from the BBO5 probe on the Bruker Avance 300 to the BBO5 probe on the Bruker Avance II+ 600 equals 3.433.
| Magritek Spinsolve 80 Carbon |
HCF |
1 |
17.333 |
36.500 |
59.500 |
66.167 |
102.167 |
| Bruker Avance 300 |
BBO5 |
0.058 |
1 |
2.106 |
3.433 |
3.817 |
5.894 |
| Bruker Avance III HD 400 |
BBFO5 |
0.027 |
0.475 |
1 |
1.630 |
1.813 |
2.799 |
| Bruker Avance II+ 600 |
BBO5 |
0.017 |
0.291 |
0.613 |
1 |
1.112 |
1.717 |
| Bruker Avance Neo 600 |
TBO5 |
0.015 |
0.262 |
0.552 |
0.899 |
1 |
1.544 |
| TCI5 |
0.010 |
0.170 |
0.357 |
0.582 |
0.648 |
1 |
Relative performance factors \(P\) for the spectrometers in automation for direct 15N observation. To determine a relative performance factor going from one spectrometer to another, read the former from the vertical header and the latter from the horizontal header. For example, the relative 15N performance factor \(P\) for switching from the BBFO5 probe on the Bruker Avance III HD 400 to the BBO5 probe on the Bruker Avance II+ 600 equals 3.600.
| Magritek Spinsolve 80 Carbon |
HCF |
N/A |
N/A |
N/A |
N/A |
N/A |
N/A |
| Bruker Avance 300 |
BBO5 |
N/A |
1 |
0.238 |
0.857 |
2.810 |
1.381 |
| Bruker Avance III HD 400 |
BBFO5 |
N/A |
4.200 |
1 |
3.600 |
11.800 |
5.800 |
| Bruker Avance II+ 600 |
BBO5 |
N/A |
1.167 |
0.278 |
1 |
3.278 |
1.611 |
| Bruker Avance Neo 600 |
TBO5 |
N/A |
0.356 |
0.085 |
0.305 |
1 |
0.492 |
| TCI5 |
N/A |
0.724 |
0.172 |
0.621 |
2.034 |
1 |
Relative performance factors \(P\) for the spectrometers in automation for direct 19F observation. To determine a relative performance factor going from one spectrometer to another, read the former from the vertical header and the latter from the horizontal header. For example, the relative 19F performance factor \(P\) for switching from the BBFO5 probe on the Bruker Avance III HD 400 to the TBO5 probe on the Bruker Avance Neo 600 equals 5.014.
| Magritek Spinsolve 80 Carbon |
HCF |
1 |
N/A |
96 |
N/A |
481.333 |
1333.333 |
| Bruker Avance 300 |
BBO5 |
N/A |
N/A |
N/A |
N/A |
N/A |
N/A |
| Bruker Avance III HD 400 |
BBFO5 |
0.010 |
N/A |
1 |
N/A |
5.014 |
13.889 |
| Bruker Avance II+ 600 |
BBO5 |
N/A |
N/A |
N/A |
N/A |
N/A |
N/A |
| Bruker Avance Neo 600 |
TBO5 |
0.002 |
N/A |
0.199 |
N/A |
1 |
2.770 |
| TCI5 |
0.001 |
N/A |
0.072 |
N/A |
0.361 |
1 |
Relative performance factors \(P\) for the spectrometers in automation for direct 29Si observation. To determine a relative performance factor going from one spectrometer to another, read the former from the vertical header and the latter from the horizontal header. For example, the relative 29Si performance factor \(P\) for switching from the BBFO5 probe on the Bruker Avance III HD 400 to the BBO5 probe on the Bruker Avance II+ 600 equals 1.545.
| Magritek Spinsolve 80 Carbon |
HCF |
N/A |
N/A |
N/A |
N/A |
N/A |
N/A |
| Bruker Avance 300 |
BBO5 |
N/A |
1 |
2.106 |
3.255 |
3.766 |
N/A |
| Bruker Avance III HD 400 |
BBFO5 |
N/A |
0.475 |
1 |
1.545 |
1.788 |
N/A |
| Bruker Avance II+ 600 |
BBO5 |
N/A |
0.307 |
0.647 |
1 |
1.157 |
N/A |
| Bruker Avance Neo 600 |
TBO5 |
N/A |
0.266 |
0.559 |
0.864 |
1 |
N/A |
| TCI5 |
N/A |
N/A |
N/A |
N/A |
N/A |
N/A |
Relative performance factors \(P\) for the spectrometers in automation for direct 29Si observation. To determine a relative performance factor going from one spectrometer to another, read the former from the vertical header and the latter from the horizontal header. For example, the relative 31P performance factor \(P\) for switching from the BBO5 probe on the Bruker Avance II+ 600 to the TBO5 probe on the Bruker Avance Neo 600 equals 3.172.
| Magritek Spinsolve 80 Carbon |
HCF |
N/A |
N/A |
N/A |
N/A |
N/A |
N/A |
| Bruker Avance 300 |
BBO5 |
N/A |
1 |
1.076 |
0.977 |
3.099 |
N/A |
| Bruker Avance III HD 400 |
BBFO5 |
N/A |
0.929 |
1 |
0.908 |
2.879 |
N/A |
| Bruker Avance II+ 600 |
BBO5 |
N/A |
1.023 |
1.102 |
1 |
3.172 |
N/A |
| Bruker Avance Neo 600 |
TBO5 |
N/A |
0.323 |
0.347 |
0.315 |
1 |
N/A |
| TCI5 |
N/A |
N/A |
N/A |
N/A |
N/A |
N/A |
For example, if you have performed a 1H experiment on the Bruker Avance 300 with a SINO of 250 and you would perform the same experiment on the Bruker Avance II+ 600, the SINO would be 250 * 4.255 = 1063.75.
Convert SINOs for different spectrometers
Use the applet below to calculate the changes in sample concentration or number of scans when changing from one spectrometer to another. Find the performance factor for the desired conversion in the table below (note the proper nucleus), by clicking on that value in the table. Next, fill in all reference values \(NS_{ref}\), \(SINO_{ref}\) and \(C_{ref}\) and two out of three of the calculated values \(NS_{calc}\), \(SINO_{calc}\) and \(C_{calc}\). The third, blank parameter will then be calculated according to the formula below. If a parameter remains constant, you can fillin the same value for the reference and calculated field, or toggle the corresponding 'Constant' button.
\[\frac{C_{calc}^2 \times NS_{calc}}{SINO_{calc}^2} = P^2 \times \frac{C_{ref}^2 \times NS_{ref}}{SINO_{ref}^2} \]