Experiments in automation
On account of the different types of probes inserted in the spectrometers, many different experiments are possible in the NMR Core Facilities. The tables below display the various experiments currently available for each spectrometer in automation. The standard 1H and 13C experiments are listed in bold. If a specific experiment is desired that is not listed below, please contact Gert Steurs.
Bruker Avance 300
| Experiment | Pulse program | Description |
|---|---|---|
| 1D 1H (30) | zg30 | 1D 1H experiment using a 30 degree flip angle |
| 1D 1H | zg | 1D 1H experiment |
| 1D 1H (PR) | zppr | 1D 1H experiment with solvent suppression via f1 presaturation |
| 1D 13C (GD30) | zggd30 | 1D 13C experiment with gated decoupling and a 30 degree flip angle |
| 1D 13C{1H} (PG30) | zgpg30 | 1D 13C{1H} experiment with power-gated decoupling and a 30 degree flip angle |
| 1D 13C{1H} (IG30) | zgig30 | 1D 13C{1H} experiment with inverse-gated decoupling and a 30 degree flip angle |
| 1D 13C{1H} DEPT135 (SP) | deptsp135 | 1D 13C{1H} DEPT-135 experiment using shaped pulses |
| 1D 13C{1H} DEPT90 (SP) | deptsp90 | 1D 13C{1H} DEPT-90 experiment using shaped pulses |
| 1D 13C{1H} DEPT45 (SP) | deptsp45 | 1D 13C{1H} DEPT-45 experiment using shaped pulses |
| 1D 13C{1H} APT | jmod | 1D 13C{1H} APT experiment |
Bruker Avance III HD 400
| Experiment | Pulse program | Description |
|---|---|---|
| 1D 1H (30) | zg30 | 1D 1H experiment using a 30 degree flip angle |
| 1D 1H | zg | 1D 1H experiment |
| 1D 1H QNMR (30) | zg30 | 1D 1H experiment for qNMR using a 30 degree flip angle. The option exact is used during ATM and a pulse calibration is performed in the background (using the pulsecal nutation experiment on the largest peak). |
| 1D 1H (PR) | zppr | 1D 1H experiment with solvent suppression via f1 presaturation |
| 1D 1H (GPPR1D) | noesygppr1d | 1D 1H experiment with solvent suppression via NOE with gradient pulses |
| 1D 1H (ESGP) | zgesgp | 1D 1H experiment with solvent suppression via excitation sculpting with gradient pulses |
| 1D 1H{13C} (IG30ADIA)$ | zgig30 | 1D 1H{13C} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle |
| 1D 1H{19F} (IG30ADIA) (QNP)#, $ | zghfigqn30 | 1D 1H{19F} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle |
| 1D 1H{31P} (IG30ADIA)$ | zgig30 | 1D 1H{31P} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle |
| 1D 2H (LOCK) | zg2h | 1D 2H experiment using the 2H lock channel. No ATM is performed. |
| 1D 7Li (30) | zg30 | 1D 7Li experiment using a 30 degree flip angle |
| 1D 11B | zg | 1D 11B experiment |
| 1D 11B{1H} (PG) | zgpg | 1D 11B{1H} experiment with power-gated decoupling |
| 1D 13C (GD30) | zggd30 | 1D 13C experiment with gated decoupling and a 30 degree flip angle |
| 1D 13C{1H} (PG30) | zgpg30 | 1D 13C{1H} experiment with power-gated decoupling and a 30 degree flip angle |
| 1D 13C{1H} (IG30) | zgig30 | 1D 13C{1H} experiment with inverse-gated decoupling and a 30 degree flip angle |
| 1D 13C{1H} DEPT135 (SP) | deptsp135 | 1D 13C{1H} DEPT-135 experiment using shaped pulses |
| 1D 13C{1H} DEPT90 (SP) | deptsp90 | 1D 13C{1H} DEPT-90 experiment using shaped pulses |
| 1D 13C{1H} DEPT45 (SP) | deptsp45 | 1D 13C{1H} DEPT-45 experiment using shaped pulses |
| 1D 13C{1H} DEPTQ135 (GPSP) | deptqgpsp | 1D 13C{1H} DEPTQ-135 experiment using shaped pulses |
| 1D 13C{1H} APT | jmod | 1D 13C{1H} APT experiment |
| 1D 15N | zg | 1D 15N experiment |
| 1D 15N{1H} (IG) | zgig | 1D 15N{1H} experiment with inverse-gated decoupling |
| 1D 19F (QNP)# | zgflqn | 1D 19F experiment |
| 1D 19F{1H} (IG) (QNP)# | zgfhigqn.2 | 1D 19F{1H} experiment with inverse-gated decoupling |
| 1D 27Al | zg | 1D 27Al experiment |
| 1D 29Si | zg | 1D 29Si experiment |
| 1D 29Si{1H} (IG) | zgig | 1D 29Si{1H} experiment with inverse-gated decoupling |
| 1D 31P (30) | zg30 | 1D 31P experiment using a 30 degree flip angle |
| 1D 31P{1H} (PG30) | zgpg30 | 1D 31P{1H} experiment with power-gated decoupling and a 30 degree flip angle |
| 1D 31P{1H} (IG30) | zgig30 | 1D 31P{1H} experiment with inverse-gated decoupling and a 30 degree flip angle |
| 1D 35Cl | zg | 1D 35Cl experiment |
| 1D 51V | zg | 1D 51V experiment |
| 1D 119Sn (30) | zg30 | 1D 119Sn experiment using a 30 degree flip angle |
| 1D 119Sn{1H} (IG30) | zgig30 | 1D 119Sn{1H} experiment with inverse-gated decoupling and a 30 degree flip angle |
| 1D 195Pt | zg | 1D 195Pt experiment |
| 1D 195Pt{1H} | zgpg | 1D 195Pt{1H} experiment with inverse-gated decoupling |
| 2D 1H T1 (IR) | t1ir | pseudo-2D 1H T1 experiment using inversion-recovery |
| 2D 1H-1H COSY (GPPPQF) | cosygpppqf | 2D 1H-1H COSY experiment using gradient and purge pulses (magnitude mode) |
| 2D 1H-1H DQF-COSY (GPDFPHPP) | cosygpmfphpp | 2D 1H-1H DQF-COSY experiment using gradient and purge pulses (phase-sensitive) |
| 2D 1H-1H LR-COSY (GPLRPPQF) | cosygplrppqf | 2D 1H-1H LR-COSY experiment using gradient and purge pulses (magnitude mode) |
| 2D 1H-1H NOESY (GPPHPP) | noesygpphpp | 2D 1H-1H NOESY experiment using gradient and purge pulses (phase-sensitive) |
| 2D 1H-1H ROESY (PHPP.2) | roesyphpp.2 | 2D 1H-1H ROESY experiment with CW spinlock (180x 180-x) for mixing and purge pulses (phase-sensitive) |
| 2D 1H-1H DOSY (LEDBPGP2S) | ledbpgp2s | 2D 1H-1H DOSY experiment using stimulated echo and LED, bipolar gradient diffusion pulses and 2 spoil pulses (phase-sensitive). Gradient pulse (p30, δ) and diffusion time (d20, Δ) are optimized in the background using the 1D version of the sequence (highest peak is chosen). |
| 2D 1H-13C HSQC-DEPT (EDETGPSISP2.2ADIA)+ | hsqcedetgpsisp2.2 | 2D 1H-13C HSQC-DEPT experiment with sensitivity improvement, using gradient, trim, shaped and adiabatic pulses |
| 2D 1H-13C HSQC (ETGPSP.2ADIA)+ | hsqcetgpsp.2 | 2D 1H-13C HSQC experiment using gradient, trim, shaped and adiabatic pulses (phase-sensitive). |
| 2D 1H-13C HSQC (ETGPSP.3)+ | hsqcetgpsp.3 | 2D 1H-13C HSQC experiment using gradient, trim and shaped pulses (phase-sensitive). For 13C-labeled molecules only. |
| 2D 1H-13C HMBC (GPLPNDQF) | hmbcgplpndqf | 2D 1H-13C HMBC experiment using gradient pulses, low-pass filter and no decoupling during acquisitoin (magnitude mode) |
| 2D 1H-13C HMBC (ETGPL3ND) | hmbcetgpl3nd | 2D 1H-13C HMBC experiment using gradient pulses, 3-fold low-pass filter and no decoupling during acquisition (phase-sensitive) |
| 2D 1H-15N HSQC (ETGPSI2)+ | hsqcetgpsi2 | 2D 1H-15N HSQC experiment with sensitivity improvement, using gradient and trim pulses |
| 2D 1H-15N HMBC (GPNDQF) | hmbcgpndqf | 2D 1H-15N HMBC experiment using gradient pulses and no decoupling (magnitude mode) |
| 2D 1H-31P HMQC (ETGPSI.2ADIA)+ | hmqcetgpsi.2 | 2D 1H-31P HMQC experiment with sensitivity improvement, using gradient, shaped and adiatbatic pulses (phase-sensitive) |
| 2D 1H-31P HMBC (GPLPNDQF) | hmbcgplpndqf | 2D 1H-31P HMBC experiment using gradient pulses, low-pass filter and no decoupling during acquisition (magnitude mode) |
| 2D 1H-195Pt HMQC (QF)+ | hmqcqf | 2D 1H-195Pt HMQC experiment (magnitude mode) |
| 2D 19F-19F COSY (GPPPQF) | cosygpppqf | 2D 19F-19F COSY experiment using gradient and purge pulses (magnitude mode) |
| 2D 19F-19F NOESY (GPPHPP) | noesygpphpp | 2D 19F-19F NOESY experiment using gradient and purge pulses (phase-sensitive) |
| 2D 31P-31P COSY (GPQF) | cosygpqf | 2D 31P-31P COSY experiment using gradient pulses (magnitude mode) |
Bruker Avance II+ 600
| Experiment | Pulse program | Description |
|---|---|---|
| 1D 1H (30) | zg30 | 1D 1H experiment using a 30 degree flip angle |
| 1D 1H | zg | 1D 1H experiment |
| 1D 1H (PR) | zppr | 1D 1H experiment with solvent suppression via f1 presaturation |
| 1D 1H (GPPR1D) | noesygppr1d | 1D 1H experiment with solvent suppression via NOE with gradient pulses |
| 1D 1H (ESGP) | zgesgp | 1D 1H experiment with solvent suppression via excitation sculpting with gradient pulses |
| 1D 1H{13C} (IG30ADIA)$ | zgig30 | 1D 1H{13C} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle |
| 1D 1H{31P} (IG30ADIA)$ | zgig30 | 1D 1H{31P} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle |
| 1D 2H (LOCK) | zg2h | 1D 2H experiment using the 2H lock channel. No ATM is performed. |
| 1D 7Li (30) | zg30 | 1D 7Li experiment using a 30 degree flip angle |
| 1D 11B | zg | 1D 11B experiment |
| 1D 11B{1H} (PG) | zgpg | 1D 11B{1H} experiment with power-gated decoupling |
| 1D 13C (GD30) | zggd30 | 1D 13C experiment with gated decoupling and a 30 degree flip angle |
| 1D 13C{1H} (PG30) | zgpg30 | 1D 13C{1H} experiment with power-gated decoupling and a 30 degree flip angle |
| 1D 13C{1H} (IG30) | zgig30 | 1D 13C{1H} experiment with inverse-gated decoupling and a 30 degree flip angle |
| 1D 13C{1H} DEPT135 (SP) | deptsp135 | 1D 13C{1H} DEPT-135 experiment using shaped pulses |
| 1D 13C{1H} DEPT90 (SP) | deptsp90 | 1D 13C{1H} DEPT-90 experiment using shaped pulses |
| 1D 13C{1H} DEPT45 (SP) | deptsp45 | 1D 13C{1H} DEPT-45 experiment using shaped pulses |
| 1D 13C{1H} DEPTQ135 (GPSP) | deptqgpsp | 1D 13C{1H} DEPTQ-135 experiment using shaped pulses |
| 1D 15N | zg | 1D 15N experiment |
| 1D 15N{1H} (IG) | zgig | 1D 15N{1H} experiment with inverse-gated decoupling |
| 1D 27Al | zg | 1D 27Al experiment |
| 1D 29Si | zg | 1D 29Si experiment |
| 1D 29Si{1H} (IG) | zgig | 1D 29Si{1H} experiment with inverse-gated decoupling |
| 1D 31P (30) | zg30 | 1D 31P experiment using a 30 degree flip angle |
| 1D 31P{1H} (PG30) | zgpg30 | 1D 31P{1H} experiment with power-gated decoupling and a 30 degree flip angle |
| 1D 31P{1H} (IG30) | zgig30 | 1D 31P{1H} experiment with inverse-gated decoupling and a 30 degree flip angle |
| 1D 35Cl | zg | 1D 35Cl experiment |
| 1D 43Ca | zg | 1D 43Ca experiment |
| 1D 43Ca (ARING) | aring | 1D 43Ca experiment using antiring sequence |
| 1D 51V | zg | 1D 51V experiment |
| 1D 119Sn (30) | zg30 | 1D 119Sn experiment using a 30 degree flip angle |
| 1D 119Sn{1H} (IG30) | zgig30 | 1D 119Sn{1H} experiment with inverse-gated decoupling and a 30 degree flip angle |
| 1D 195Pt | zg | 1D 195Pt experiment |
| 1D 195Pt{1H} | zgpg | 1D 195Pt{1H} experiment with inverse-gated decoupling |
| 2D 1H T1 (IR) | t1ir | pseudo-2D 1H T1 experiment using inversion-recovery |
| 2D 1H-1H COSY (GPPPQF) | cosygpppqf | 2D 1H-1H COSY experiment using gradient and purge pulses (magnitude mode) |
| 2D 1H-1H DQF-COSY (GPDFPHPP) | cosygpmfphpp | 2D 1H-1H DQF-COSY experiment using gradient and purge pulses (phase-sensitive) |
| 2D 1H-1H LR-COSY (GPLRPPQF) | cosygplrppqf | 2D 1H-1H LR-COSY experiment using gradient and purge pulses (magnitude mode) |
| 2D 1H-1H NOESY (GPPHPP) | noesygpphpp | 2D 1H-1H NOESY experiment using gradient and purge pulses (phase-sensitive) |
| 2D 1H-1H ROESY (PHPP.2) | roesyphpp.2 | 2D 1H-1H ROESY experiment with CW spinlock (180x 180-x) for mixing and purge pulses (phase-sensitive) |
| 2D 1H-1H DOSY (LEDBPGP2S) | ledbpgp2s | 2D 1H-1H DOSY experiment using stimulated echo and LED, bipolar gradient diffusion pulses and 2 spoil pulses (phase-sensitive). Gradient pulse (p30, δ) and diffusion time (d20, Δ) are optimized in the background using the 1D version of the sequence (highest peak is chosen). |
| 2D 1H-13C HSQC-DEPT (EDETGPSISP2.2ADIA)+ | hsqcedetgpsisp2.2 | 2D 1H-13C HSQC-DEPT experiment with sensitivity improvement, using gradient, trim, shaped and adiabatic pulses |
| 2D 1H-13C HSQC (ETGPSP.2ADIA)+ | hsqcetgpsp.2 | 2D 1H-13C HSQC experiment using gradient, trim, shaped and adiabatic pulses (phase-sensitive). |
| 2D 1H-13C HMBC (GPLPNDQF) | hmbcgplpndqf | 2D 1H-13C HMBC experiment using gradient pulses, low-pass filter and no decoupling during acquisitoin (magnitude mode) |
| 2D 1H-13C HMBC (ETGPL3ND) | hmbcetgpl3nd | 2D 1H-13C HMBC experiment using gradient pulses, 3-fold low-pass filter and no decoupling during acquisition (phase-sensitive) |
| 2D 1H-15N HSQC (ETGPSI2)+ | hsqcetgpsi2 | 2D 1H-15N HSQC experiment with sensitivity improvement, using gradient and trim pulses |
| 2D 1H-15N HMBC (GPNDQF) | hmbcgpndqf | 2D 1H-15N HMBC experiment using gradient pulses and no decoupling (magnitude mode) |
| 2D 1H-31P HMQC (ETGPSI.2ADIA)+ | hmqcetgpsi.2 | 2D 1H-31P HMQC experiment with sensitivity improvement, using gradient, shaped and adiatbatic pulses (phase-sensitive) |
| 2D 1H-31P HMBC (GPLPNDQF) | hmbcgplpndqf | 2D 1H-31P HMBC experiment using gradient pulses, low-pass filter and no decoupling during acquisition (magnitude mode) |
| 2D 1H-195Pt HMQC (QF)+ | hmqcqf | 2D 1H-195Pt HMQC experiment (magnitude mode) |
| 2D 31P-31P COSY (GPQF) | cosygpqf | 2D 31P-31P COSY experiment using gradient pulses (magnitude mode) |
Bruker Avance Neo 600
| Experiment | Pulse program | Description | Probe |
|---|---|---|---|
| 1D 1H (30) | zg30 | 1D 1H experiment using a 30 degree flip angle | TCI5, TBO5 |
| 1D 1H | zg | 1D 1H experiment | TCI5, TBO5 |
| 1D 1H (PR) | zppr | 1D 1H experiment with solvent suppression via f1 presaturation | TCI5, TBO5 |
| 1D 1H (GPPR1D) | noesygppr1d | 1D 1H experiment with solvent suppression via NOE with gradient pulses | TCI5, TBO5 |
| 1D 1H (ESGP) | zgesgp | 1D 1H experiment with solvent suppression via excitation sculpting with gradient pulses | TCI5, TBO5 |
| 1D 1H{13C} (IG30ADIA)$ | zgig30 | 1D 1H{13C} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle | TCI5, TBO5 |
| 1D 1H{19F} (IG30ADIA)$ | zgig30 | 1D 1H{19F} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle | TBO5 |
| 1D 1H{31P} (IG30ADIA)$ | zgig30 | 1D 1H{31P} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle | TBO5 |
| 1D 1H{13C,19F} (IG30ADIA)$ | zgfbig | 1D 1H{13C,19F} experiment with inverse-gated adiabatic decoupling on both channels | TBO5 |
| 1D 2H (LOCK) | zg2h | 1D 2H experiment using the 2H lock channel. No ATM is performed. | TCI5, TBO5 |
| 1D 7Li (30) | zg30 | 1D 7Li experiment using a 30 degree flip angle | TBO5 |
| 1D 11B | zg | 1D 11B experiment | TBO5 |
| 1D 11B{1H} (PG) | zgpg | 1D 11B{1H} experiment with power-gated decoupling | TBO5 |
| 1D 13C (GD30) | zggd30 | 1D 13C experiment with gated decoupling and a 30 degree flip angle | TCI5, TBO5 |
| 1D 13C{1H} (PG30) | zgpg30 | 1D 13C{1H} experiment with power-gated decoupling and a 30 degree flip angle | TCI5, TBO5 |
| 1D 13C{1H} (IG30) | zgig30 | 1D 13C{1H} experiment with inverse-gated decoupling and a 30 degree flip angle | TCI5, TBO5 |
| 1D 13C{1H} DEPT135 (SP) | deptsp135 | 1D 13C{1H} DEPT-135 experiment using shaped pulses | TCI5, TBO5 |
| 1D 13C{1H} DEPT90 (SP) | deptsp90 | 1D 13C{1H} DEPT-90 experiment using shaped pulses | TCI5, TBO5 |
| 1D 13C{1H} DEPT45 (SP) | deptsp45 | 1D 13C{1H} DEPT-45 experiment using shaped pulses | TCI5, TBO5 |
| 1D 13C{1H} DEPTQ135 (GPSP) | deptqgpsp | 1D 13C{1H} DEPTQ-135 experiment using shaped pulses | TCI5, TBO5 |
| 1D 15N | zg | 1D 15N experiment | TCI5, TBO5 |
| 1D 15N{1H} (IG) | zgig | 1D 15N{1H} experiment with inverse-gated decoupling | TCI5, TBO5 |
| 1D 19F | zg | 1D 19F experiment | TCI5, TBO5 |
| 1D 19F{1H} (IG) | zgig | 1D 19F{1H} experiment with inverse-gated decoupling | TBO5 |
| 1D 27Al | zg | 1D 27Al experiment | TBO5 |
| 1D 29Si | zg | 1D 29Si experiment | TBO5 |
| 1D 29Si{1H} (IG) | zgig | 1D 29Si{1H} experiment with inverse-gated decoupling | TBO5 |
| 1D 31P (30) | zg30 | 1D 31P experiment using a 30 degree flip angle | TBO5 |
| 1D 31P{1H} (PG30) | zgpg30 | 1D 31P{1H} experiment with power-gated decoupling and a 30 degree flip angle | TBO5 |
| 1D 31P{1H} (IG30) | zgig30 | 1D 31P{1H} experiment with inverse-gated decoupling and a 30 degree flip angle | TBO5 |
| 1D 35Cl | zg | 1D 35Cl experiment | TBO5 |
| 1D 43Ca | zg | 1D 43Ca experiment | TBO5 |
| 1D 43Ca (ARING) | aring | 1D 43Ca experiment using antiring sequence | TBO5 |
| 1D 51V | zg | 1D 51V experiment | TBO5 |
| 1D 119Sn (30) | zg30 | 1D 119Sn experiment using a 30 degree flip angle | TBO5 |
| 1D 119Sn{1H} (IG30) | zgig30 | 1D 119Sn{1H} experiment with inverse-gated decoupling and a 30 degree flip angle | TBO5 |
| 1D 195Pt | zg | 1D 195Pt experiment | TBO5 |
| 1D 195Pt{1H} | zgpg | 1D 195Pt{1H} experiment with inverse-gated decoupling | TBO5 |
| 2D 1H T1 (IR) | t1ir | pseudo-2D 1H T1 experiment using inversion-recovery | TCI5, TBO5 |
| 2D 1H-1H COSY (GPPPQF) | cosygpppqf | 2D 1H-1H COSY experiment using gradient and purge pulses (magnitude mode) | TCI5, TBO5 |
| 2D 1H-1H DQF-COSY (GPDFPHPP) | cosygpmfphpp | 2D 1H-1H DQF-COSY experiment using gradient and purge pulses (phase-sensitive) | TCI5, TBO5 |
| 2D 1H-1H LR-COSY (GPLRPPQF) | cosygplrppqf | 2D 1H-1H LR-COSY experiment using gradient and purge pulses (magnitude mode) | TCI5, TBO5 |
| 2D 1H-1H NOESY (GPPHPP) | noesygpphpp | 2D 1H-1H NOESY experiment using gradient and purge pulses (phase-sensitive) | TCI5, TBO5 |
| 2D 1H-1H ROESY (PHPP.2) | roesyphpp.2 | 2D 1H-1H ROESY experiment with CW spinlock (180x 180-x) for mixing and purge pulses (phase-sensitive) | TCI5, TBO5 |
| 2D 1H-1H DOSY (LEDBPGP2S) | ledbpgp2s | 2D 1H-1H DOSY experiment using stimulated echo and LED, bipolar gradient diffusion pulses and 2 spoil pulses (phase-sensitive). Gradient pulse (p30, δ) and diffusion time (d20, Δ) are optimized in the background using the 1D version of the sequence (highest peak is chosen). | TCI5, TBO5 |
| 2D 1H-13C HSQC-DEPT (EDETGPSISP2.2ADIA)+ | hsqcedetgpsisp2.2 | 2D 1H-13C HSQC-DEPT experiment with sensitivity improvement, using gradient, trim, shaped and adiabatic pulses | TCI5, TBO5 |
| 2D 1H-13C HSQC (ETGPSP.2ADIA)+ | hsqcetgpsp.2 | 2D 1H-13C HSQC experiment using gradient, trim, shaped and adiabatic pulses (phase-sensitive). | TCI5, TBO5 |
| 2D 1H-13C HMBC (GPLPNDQF) | hmbcgplpndqf | 2D 1H-13C HMBC experiment using gradient pulses, low-pass filter and no decoupling during acquisitoin (magnitude mode) | TCI5, TBO5 |
| 2D 1H-13C HMBC (ETGPL3ND) | hmbcetgpl3nd | 2D 1H-13C HMBC experiment using gradient pulses, 3-fold low-pass filter and no decoupling during acquisition (phase-sensitive) | TCI5, TBO5 |
| 2D 1H-15N HSQC (ETGPSI2)+ | hsqcetgpsi2 | 2D 1H-15N HSQC experiment with sensitivity improvement, using gradient and trim pulses | TCI5, TBO5 |
| 2D 1H-15N HMBC (GPNDQF) | hmbcgpndqf | 2D 1H-15N HMBC experiment using gradient pulses and no decoupling (magnitude mode) | TCI5, TBO5 |
| 2D 1H-31P HMQC (ETGPSI.2ADIA)+ | hmqcetgpsi.2 | 2D 1H-31P HMQC experiment with sensitivity improvement, using gradient, shaped and adiatbatic pulses (phase-sensitive) | TBO5 |
| 2D 1H-31P HMBC (GPLPNDQF) | hmbcgplpndqf | 2D 1H-31P HMBC experiment using gradient pulses, low-pass filter and no decoupling during acquisition (magnitude mode) | TBO5 |
| 2D 1H-195Pt HMQC (QF)+ | hmqcqf | 2D 1H-195Pt HMQC experiment (magnitude mode) | TBO5 |
| 2D 19F-13C HMQC (ETGPSI.2)+ | hmqcetgpsi.2 | 2D 19F-13C HMQC experiment with sensitivity improvement, using gradient and shaped pulses (phase-sensitive) | TCI5, TBO5 |
| 2D 19F-13C HMBC (GPLPNDQF) | hmbcgplpndqf | 2D 19F-13C HMBC experiment using gradient pulses, low-pass filter and no decoupling during acquisition (magnitude mode) | TCI5, TBO5 |
| 2D 19F-19F COSY (GPPPQF) | cosygpppqf | 2D 19F-19F COSY experiment using gradient and purge pulses (magnitude mode) | TCI5, TBO5 |
| 2D 19F-19F NOESY (GPPHPP) | noesygpphpp | 2D 19F-19F NOESY experiment using gradient and purge pulses (phase-sensitive) | TCI5, TBO5 |
| 2D 31P-31P COSY (GPQF) | cosygpqf | 2D 31P-31P COSY experiment using gradient pulses (magnitude mode) | TBO5 |
Footnotes
#
Pulse sequence adapted for use with a spectrometer equipped with a QNP switch. The essence of the sequence is exactly the same as its non-QNP counterpart.
$
Due to the (adiabatic) broadband-decoupling during the acquisition time (AQ), the latter is restricted to a maximum of 0.4 sec for the Bruker Avance III HD 400 and the Bruker Avance Neo 600 and 0.25 sec for the Bruker Avance II+ 600 with a total decoupling power level of maximum 5 W. Note that the former also affects the maximum TD. A minimum D1 relaxation time of 5 sec is required.
+
Due to the broadband-decoupling during the acquisition time (AQ (F2)), the latter is restricted to a maximum of 0.2 sec. Note that this also affects the maximum TD (F2).
For an explanation of the two-character pulse program abbreviations used, please refer to the list of experiment abbreviations.