Experiments in automation

On account of the different types of probes inserted in the spectrometers, many different experiments are possible in the NMR Core Facilities. The tables below display the various experiments currently available for each spectrometer in automation. The standard 1H and 13C experiments are listed in bold. If a specific experiment is desired that is not listed below, please contact Gert Steurs.

Bruker Avance 300

Experiment Pulse program Description
1D 1H (30) zg30 1D 1H experiment using a 30 degree flip angle
1D 1H zg 1D 1H experiment
1D 1H (PR) zppr 1D 1H experiment with solvent suppression via f1 presaturation
1D 13C (GD30) zggd30 1D 13C experiment with gated decoupling and a 30 degree flip angle
1D 13C{1H} (PG30) zgpg30 1D 13C{1H} experiment with power-gated decoupling and a 30 degree flip angle
1D 13C{1H} (IG30) zgig30 1D 13C{1H} experiment with inverse-gated decoupling and a 30 degree flip angle
1D 13C{1H} DEPT135 (SP) deptsp135 1D 13C{1H} DEPT-135 experiment using shaped pulses
1D 13C{1H} DEPT90 (SP) deptsp90 1D 13C{1H} DEPT-90 experiment using shaped pulses
1D 13C{1H} DEPT45 (SP) deptsp45 1D 13C{1H} DEPT-45 experiment using shaped pulses
1D 13C{1H} APT jmod 1D 13C{1H} APT experiment

Bruker Avance III HD 400

Experiment Pulse program Description
1D 1H (30) zg30 1D 1H experiment using a 30 degree flip angle
1D 1H zg 1D 1H experiment
1D 1H QNMR (30) zg30 1D 1H experiment for qNMR using a 30 degree flip angle. The option exact is used during ATM and a pulse calibration is performed in the background (using the pulsecal nutation experiment on the largest peak).
1D 1H (PR) zppr 1D 1H experiment with solvent suppression via f1 presaturation
1D 1H (GPPR1D) noesygppr1d 1D 1H experiment with solvent suppression via NOE with gradient pulses
1D 1H (ESGP) zgesgp 1D 1H experiment with solvent suppression via excitation sculpting with gradient pulses
1D 1H{13C} (IG30ADIA)$ zgig30 1D 1H{13C} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle
1D 1H{19F} (IG30ADIA) (QNP)#, $ zghfigqn30 1D 1H{19F} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle
1D 1H{31P} (IG30ADIA)$ zgig30 1D 1H{31P} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle
1D 2H (LOCK) zg2h 1D 2H experiment using the 2H lock channel. No ATM is performed.
1D 7Li (30) zg30 1D 7Li experiment using a 30 degree flip angle
1D 11B zg 1D 11B experiment
1D 11B{1H} (PG) zgpg 1D 11B{1H} experiment with power-gated decoupling
1D 13C (GD30) zggd30 1D 13C experiment with gated decoupling and a 30 degree flip angle
1D 13C{1H} (PG30) zgpg30 1D 13C{1H} experiment with power-gated decoupling and a 30 degree flip angle
1D 13C{1H} (IG30) zgig30 1D 13C{1H} experiment with inverse-gated decoupling and a 30 degree flip angle
1D 13C{1H} DEPT135 (SP) deptsp135 1D 13C{1H} DEPT-135 experiment using shaped pulses
1D 13C{1H} DEPT90 (SP) deptsp90 1D 13C{1H} DEPT-90 experiment using shaped pulses
1D 13C{1H} DEPT45 (SP) deptsp45 1D 13C{1H} DEPT-45 experiment using shaped pulses
1D 13C{1H} DEPTQ135 (GPSP) deptqgpsp 1D 13C{1H} DEPTQ-135 experiment using shaped pulses
1D 13C{1H} APT jmod 1D 13C{1H} APT experiment
1D 15N zg 1D 15N experiment
1D 15N{1H} (IG) zgig 1D 15N{1H} experiment with inverse-gated decoupling
1D 19F (QNP)# zgflqn 1D 19F experiment
1D 19F{1H} (IG) (QNP)# zgfhigqn.2 1D 19F{1H} experiment with inverse-gated decoupling
1D 27Al zg 1D 27Al experiment
1D 29Si zg 1D 29Si experiment
1D 29Si{1H} (IG) zgig 1D 29Si{1H} experiment with inverse-gated decoupling
1D 31P (30) zg30 1D 31P experiment using a 30 degree flip angle
1D 31P{1H} (PG30) zgpg30 1D 31P{1H} experiment with power-gated decoupling and a 30 degree flip angle
1D 31P{1H} (IG30) zgig30 1D 31P{1H} experiment with inverse-gated decoupling and a 30 degree flip angle
1D 35Cl zg 1D 35Cl experiment
1D 51V zg 1D 51V experiment
1D 119Sn (30) zg30 1D 119Sn experiment using a 30 degree flip angle
1D 119Sn{1H} (IG30) zgig30 1D 119Sn{1H} experiment with inverse-gated decoupling and a 30 degree flip angle
1D 195Pt zg 1D 195Pt experiment
1D 195Pt{1H} zgpg 1D 195Pt{1H} experiment with inverse-gated decoupling
2D 1H T1 (IR) t1ir pseudo-2D 1H T1 experiment using inversion-recovery
2D 1H-1H COSY (GPPPQF) cosygpppqf 2D 1H-1H COSY experiment using gradient and purge pulses (magnitude mode)
2D 1H-1H DQF-COSY (GPDFPHPP) cosygpmfphpp 2D 1H-1H DQF-COSY experiment using gradient and purge pulses (phase-sensitive)
2D 1H-1H LR-COSY (GPLRPPQF) cosygplrppqf 2D 1H-1H LR-COSY experiment using gradient and purge pulses (magnitude mode)
2D 1H-1H NOESY (GPPHPP) noesygpphpp 2D 1H-1H NOESY experiment using gradient and purge pulses (phase-sensitive)
2D 1H-1H ROESY (PHPP.2) roesyphpp.2 2D 1H-1H ROESY experiment with CW spinlock (180x 180-x) for mixing and purge pulses (phase-sensitive)
2D 1H-1H DOSY (LEDBPGP2S) ledbpgp2s 2D 1H-1H DOSY experiment using stimulated echo and LED, bipolar gradient diffusion pulses and 2 spoil pulses (phase-sensitive). Gradient pulse (p30, δ) and diffusion time (d20, Δ) are optimized in the background using the 1D version of the sequence (highest peak is chosen).
2D 1H-13C HSQC-DEPT (EDETGPSISP2.2ADIA)+ hsqcedetgpsisp2.2 2D 1H-13C HSQC-DEPT experiment with sensitivity improvement, using gradient, trim, shaped and adiabatic pulses
2D 1H-13C HSQC (ETGPSP.2ADIA)+ hsqcetgpsp.2 2D 1H-13C HSQC experiment using gradient, trim, shaped and adiabatic pulses (phase-sensitive).
2D 1H-13C HSQC (ETGPSP.3)+ hsqcetgpsp.3 2D 1H-13C HSQC experiment using gradient, trim and shaped pulses (phase-sensitive). For 13C-labeled molecules only.
2D 1H-13C HMBC (GPLPNDQF) hmbcgplpndqf 2D 1H-13C HMBC experiment using gradient pulses, low-pass filter and no decoupling during acquisitoin (magnitude mode)
2D 1H-13C HMBC (ETGPL3ND) hmbcetgpl3nd 2D 1H-13C HMBC experiment using gradient pulses, 3-fold low-pass filter and no decoupling during acquisition (phase-sensitive)
2D 1H-15N HSQC (ETGPSI2)+ hsqcetgpsi2 2D 1H-15N HSQC experiment with sensitivity improvement, using gradient and trim pulses
2D 1H-15N HMBC (GPNDQF) hmbcgpndqf 2D 1H-15N HMBC experiment using gradient pulses and no decoupling (magnitude mode)
2D 1H-31P HMQC (ETGPSI.2ADIA)+ hmqcetgpsi.2 2D 1H-31P HMQC experiment with sensitivity improvement, using gradient, shaped and adiatbatic pulses (phase-sensitive)
2D 1H-31P HMBC (GPLPNDQF) hmbcgplpndqf 2D 1H-31P HMBC experiment using gradient pulses, low-pass filter and no decoupling during acquisition (magnitude mode)
2D 1H-195Pt HMQC (QF)+ hmqcqf 2D 1H-195Pt HMQC experiment (magnitude mode)
2D 19F-19F COSY (GPPPQF) cosygpppqf 2D 19F-19F COSY experiment using gradient and purge pulses (magnitude mode)
2D 19F-19F NOESY (GPPHPP) noesygpphpp 2D 19F-19F NOESY experiment using gradient and purge pulses (phase-sensitive)
2D 31P-31P COSY (GPQF) cosygpqf 2D 31P-31P COSY experiment using gradient pulses (magnitude mode)

Bruker Avance II+ 600

Experiment Pulse program Description
1D 1H (30) zg30 1D 1H experiment using a 30 degree flip angle
1D 1H zg 1D 1H experiment
1D 1H (PR) zppr 1D 1H experiment with solvent suppression via f1 presaturation
1D 1H (GPPR1D) noesygppr1d 1D 1H experiment with solvent suppression via NOE with gradient pulses
1D 1H (ESGP) zgesgp 1D 1H experiment with solvent suppression via excitation sculpting with gradient pulses
1D 1H{13C} (IG30ADIA)$ zgig30 1D 1H{13C} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle
1D 1H{31P} (IG30ADIA)$ zgig30 1D 1H{31P} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle
1D 2H (LOCK) zg2h 1D 2H experiment using the 2H lock channel. No ATM is performed.
1D 7Li (30) zg30 1D 7Li experiment using a 30 degree flip angle
1D 11B zg 1D 11B experiment
1D 11B{1H} (PG) zgpg 1D 11B{1H} experiment with power-gated decoupling
1D 13C (GD30) zggd30 1D 13C experiment with gated decoupling and a 30 degree flip angle
1D 13C{1H} (PG30) zgpg30 1D 13C{1H} experiment with power-gated decoupling and a 30 degree flip angle
1D 13C{1H} (IG30) zgig30 1D 13C{1H} experiment with inverse-gated decoupling and a 30 degree flip angle
1D 13C{1H} DEPT135 (SP) deptsp135 1D 13C{1H} DEPT-135 experiment using shaped pulses
1D 13C{1H} DEPT90 (SP) deptsp90 1D 13C{1H} DEPT-90 experiment using shaped pulses
1D 13C{1H} DEPT45 (SP) deptsp45 1D 13C{1H} DEPT-45 experiment using shaped pulses
1D 13C{1H} DEPTQ135 (GPSP) deptqgpsp 1D 13C{1H} DEPTQ-135 experiment using shaped pulses
1D 15N zg 1D 15N experiment
1D 15N{1H} (IG) zgig 1D 15N{1H} experiment with inverse-gated decoupling
1D 27Al zg 1D 27Al experiment
1D 29Si zg 1D 29Si experiment
1D 29Si{1H} (IG) zgig 1D 29Si{1H} experiment with inverse-gated decoupling
1D 31P (30) zg30 1D 31P experiment using a 30 degree flip angle
1D 31P{1H} (PG30) zgpg30 1D 31P{1H} experiment with power-gated decoupling and a 30 degree flip angle
1D 31P{1H} (IG30) zgig30 1D 31P{1H} experiment with inverse-gated decoupling and a 30 degree flip angle
1D 35Cl zg 1D 35Cl experiment
1D 43Ca zg 1D 43Ca experiment
1D 43Ca (ARING) aring 1D 43Ca experiment using antiring sequence
1D 51V zg 1D 51V experiment
1D 119Sn (30) zg30 1D 119Sn experiment using a 30 degree flip angle
1D 119Sn{1H} (IG30) zgig30 1D 119Sn{1H} experiment with inverse-gated decoupling and a 30 degree flip angle
1D 195Pt zg 1D 195Pt experiment
1D 195Pt{1H} zgpg 1D 195Pt{1H} experiment with inverse-gated decoupling
2D 1H T1 (IR) t1ir pseudo-2D 1H T1 experiment using inversion-recovery
2D 1H-1H COSY (GPPPQF) cosygpppqf 2D 1H-1H COSY experiment using gradient and purge pulses (magnitude mode)
2D 1H-1H DQF-COSY (GPDFPHPP) cosygpmfphpp 2D 1H-1H DQF-COSY experiment using gradient and purge pulses (phase-sensitive)
2D 1H-1H LR-COSY (GPLRPPQF) cosygplrppqf 2D 1H-1H LR-COSY experiment using gradient and purge pulses (magnitude mode)
2D 1H-1H NOESY (GPPHPP) noesygpphpp 2D 1H-1H NOESY experiment using gradient and purge pulses (phase-sensitive)
2D 1H-1H ROESY (PHPP.2) roesyphpp.2 2D 1H-1H ROESY experiment with CW spinlock (180x 180-x) for mixing and purge pulses (phase-sensitive)
2D 1H-1H DOSY (LEDBPGP2S) ledbpgp2s 2D 1H-1H DOSY experiment using stimulated echo and LED, bipolar gradient diffusion pulses and 2 spoil pulses (phase-sensitive). Gradient pulse (p30, δ) and diffusion time (d20, Δ) are optimized in the background using the 1D version of the sequence (highest peak is chosen).
2D 1H-13C HSQC-DEPT (EDETGPSISP2.2ADIA)+ hsqcedetgpsisp2.2 2D 1H-13C HSQC-DEPT experiment with sensitivity improvement, using gradient, trim, shaped and adiabatic pulses
2D 1H-13C HSQC (ETGPSP.2ADIA)+ hsqcetgpsp.2 2D 1H-13C HSQC experiment using gradient, trim, shaped and adiabatic pulses (phase-sensitive).
2D 1H-13C HMBC (GPLPNDQF) hmbcgplpndqf 2D 1H-13C HMBC experiment using gradient pulses, low-pass filter and no decoupling during acquisitoin (magnitude mode)
2D 1H-13C HMBC (ETGPL3ND) hmbcetgpl3nd 2D 1H-13C HMBC experiment using gradient pulses, 3-fold low-pass filter and no decoupling during acquisition (phase-sensitive)
2D 1H-15N HSQC (ETGPSI2)+ hsqcetgpsi2 2D 1H-15N HSQC experiment with sensitivity improvement, using gradient and trim pulses
2D 1H-15N HMBC (GPNDQF) hmbcgpndqf 2D 1H-15N HMBC experiment using gradient pulses and no decoupling (magnitude mode)
2D 1H-31P HMQC (ETGPSI.2ADIA)+ hmqcetgpsi.2 2D 1H-31P HMQC experiment with sensitivity improvement, using gradient, shaped and adiatbatic pulses (phase-sensitive)
2D 1H-31P HMBC (GPLPNDQF) hmbcgplpndqf 2D 1H-31P HMBC experiment using gradient pulses, low-pass filter and no decoupling during acquisition (magnitude mode)
2D 1H-195Pt HMQC (QF)+ hmqcqf 2D 1H-195Pt HMQC experiment (magnitude mode)
2D 31P-31P COSY (GPQF) cosygpqf 2D 31P-31P COSY experiment using gradient pulses (magnitude mode)

Bruker Avance Neo 600

Experiment Pulse program Description Probe
1D 1H (30) zg30 1D 1H experiment using a 30 degree flip angle TCI5, TBO5
1D 1H zg 1D 1H experiment TCI5, TBO5
1D 1H (PR) zppr 1D 1H experiment with solvent suppression via f1 presaturation TCI5, TBO5
1D 1H (GPPR1D) noesygppr1d 1D 1H experiment with solvent suppression via NOE with gradient pulses TCI5, TBO5
1D 1H (ESGP) zgesgp 1D 1H experiment with solvent suppression via excitation sculpting with gradient pulses TCI5, TBO5
1D 1H{13C} (IG30ADIA)$ zgig30 1D 1H{13C} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle TCI5, TBO5
1D 1H{19F} (IG30ADIA)$ zgig30 1D 1H{19F} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle TBO5
1D 1H{31P} (IG30ADIA)$ zgig30 1D 1H{31P} experiment with inverse-gated adiabatic decoupling and a 30 degree flip angle TBO5
1D 1H{13C,19F} (IG30ADIA)$ zgfbig 1D 1H{13C,19F} experiment with inverse-gated adiabatic decoupling on both channels TBO5
1D 2H (LOCK) zg2h 1D 2H experiment using the 2H lock channel. No ATM is performed. TCI5, TBO5
1D 7Li (30) zg30 1D 7Li experiment using a 30 degree flip angle TBO5
1D 11B zg 1D 11B experiment TBO5
1D 11B{1H} (PG) zgpg 1D 11B{1H} experiment with power-gated decoupling TBO5
1D 13C (GD30) zggd30 1D 13C experiment with gated decoupling and a 30 degree flip angle TCI5, TBO5
1D 13C{1H} (PG30) zgpg30 1D 13C{1H} experiment with power-gated decoupling and a 30 degree flip angle TCI5, TBO5
1D 13C{1H} (IG30) zgig30 1D 13C{1H} experiment with inverse-gated decoupling and a 30 degree flip angle TCI5, TBO5
1D 13C{1H} DEPT135 (SP) deptsp135 1D 13C{1H} DEPT-135 experiment using shaped pulses TCI5, TBO5
1D 13C{1H} DEPT90 (SP) deptsp90 1D 13C{1H} DEPT-90 experiment using shaped pulses TCI5, TBO5
1D 13C{1H} DEPT45 (SP) deptsp45 1D 13C{1H} DEPT-45 experiment using shaped pulses TCI5, TBO5
1D 13C{1H} DEPTQ135 (GPSP) deptqgpsp 1D 13C{1H} DEPTQ-135 experiment using shaped pulses TCI5, TBO5
1D 15N zg 1D 15N experiment TCI5, TBO5
1D 15N{1H} (IG) zgig 1D 15N{1H} experiment with inverse-gated decoupling TCI5, TBO5
1D 19F zg 1D 19F experiment TCI5, TBO5
1D 19F{1H} (IG) zgig 1D 19F{1H} experiment with inverse-gated decoupling TBO5
1D 27Al zg 1D 27Al experiment TBO5
1D 29Si zg 1D 29Si experiment TBO5
1D 29Si{1H} (IG) zgig 1D 29Si{1H} experiment with inverse-gated decoupling TBO5
1D 31P (30) zg30 1D 31P experiment using a 30 degree flip angle TBO5
1D 31P{1H} (PG30) zgpg30 1D 31P{1H} experiment with power-gated decoupling and a 30 degree flip angle TBO5
1D 31P{1H} (IG30) zgig30 1D 31P{1H} experiment with inverse-gated decoupling and a 30 degree flip angle TBO5
1D 35Cl zg 1D 35Cl experiment TBO5
1D 43Ca zg 1D 43Ca experiment TBO5
1D 43Ca (ARING) aring 1D 43Ca experiment using antiring sequence TBO5
1D 51V zg 1D 51V experiment TBO5
1D 119Sn (30) zg30 1D 119Sn experiment using a 30 degree flip angle TBO5
1D 119Sn{1H} (IG30) zgig30 1D 119Sn{1H} experiment with inverse-gated decoupling and a 30 degree flip angle TBO5
1D 195Pt zg 1D 195Pt experiment TBO5
1D 195Pt{1H} zgpg 1D 195Pt{1H} experiment with inverse-gated decoupling TBO5
2D 1H T1 (IR) t1ir pseudo-2D 1H T1 experiment using inversion-recovery TCI5, TBO5
2D 1H-1H COSY (GPPPQF) cosygpppqf 2D 1H-1H COSY experiment using gradient and purge pulses (magnitude mode) TCI5, TBO5
2D 1H-1H DQF-COSY (GPDFPHPP) cosygpmfphpp 2D 1H-1H DQF-COSY experiment using gradient and purge pulses (phase-sensitive) TCI5, TBO5
2D 1H-1H LR-COSY (GPLRPPQF) cosygplrppqf 2D 1H-1H LR-COSY experiment using gradient and purge pulses (magnitude mode) TCI5, TBO5
2D 1H-1H NOESY (GPPHPP) noesygpphpp 2D 1H-1H NOESY experiment using gradient and purge pulses (phase-sensitive) TCI5, TBO5
2D 1H-1H ROESY (PHPP.2) roesyphpp.2 2D 1H-1H ROESY experiment with CW spinlock (180x 180-x) for mixing and purge pulses (phase-sensitive) TCI5, TBO5
2D 1H-1H DOSY (LEDBPGP2S) ledbpgp2s 2D 1H-1H DOSY experiment using stimulated echo and LED, bipolar gradient diffusion pulses and 2 spoil pulses (phase-sensitive). Gradient pulse (p30, δ) and diffusion time (d20, Δ) are optimized in the background using the 1D version of the sequence (highest peak is chosen). TCI5, TBO5
2D 1H-13C HSQC-DEPT (EDETGPSISP2.2ADIA)+ hsqcedetgpsisp2.2 2D 1H-13C HSQC-DEPT experiment with sensitivity improvement, using gradient, trim, shaped and adiabatic pulses TCI5, TBO5
2D 1H-13C HSQC (ETGPSP.2ADIA)+ hsqcetgpsp.2 2D 1H-13C HSQC experiment using gradient, trim, shaped and adiabatic pulses (phase-sensitive). TCI5, TBO5
2D 1H-13C HMBC (GPLPNDQF) hmbcgplpndqf 2D 1H-13C HMBC experiment using gradient pulses, low-pass filter and no decoupling during acquisitoin (magnitude mode) TCI5, TBO5
2D 1H-13C HMBC (ETGPL3ND) hmbcetgpl3nd 2D 1H-13C HMBC experiment using gradient pulses, 3-fold low-pass filter and no decoupling during acquisition (phase-sensitive) TCI5, TBO5
2D 1H-15N HSQC (ETGPSI2)+ hsqcetgpsi2 2D 1H-15N HSQC experiment with sensitivity improvement, using gradient and trim pulses TCI5, TBO5
2D 1H-15N HMBC (GPNDQF) hmbcgpndqf 2D 1H-15N HMBC experiment using gradient pulses and no decoupling (magnitude mode) TCI5, TBO5
2D 1H-31P HMQC (ETGPSI.2ADIA)+ hmqcetgpsi.2 2D 1H-31P HMQC experiment with sensitivity improvement, using gradient, shaped and adiatbatic pulses (phase-sensitive) TBO5
2D 1H-31P HMBC (GPLPNDQF) hmbcgplpndqf 2D 1H-31P HMBC experiment using gradient pulses, low-pass filter and no decoupling during acquisition (magnitude mode) TBO5
2D 1H-195Pt HMQC (QF)+ hmqcqf 2D 1H-195Pt HMQC experiment (magnitude mode) TBO5
2D 19F-13C HMQC (ETGPSI.2)+ hmqcetgpsi.2 2D 19F-13C HMQC experiment with sensitivity improvement, using gradient and shaped pulses (phase-sensitive) TCI5, TBO5
2D 19F-13C HMBC (GPLPNDQF) hmbcgplpndqf 2D 19F-13C HMBC experiment using gradient pulses, low-pass filter and no decoupling during acquisition (magnitude mode) TCI5, TBO5
2D 19F-19F COSY (GPPPQF) cosygpppqf 2D 19F-19F COSY experiment using gradient and purge pulses (magnitude mode) TCI5, TBO5
2D 19F-19F NOESY (GPPHPP) noesygpphpp 2D 19F-19F NOESY experiment using gradient and purge pulses (phase-sensitive) TCI5, TBO5
2D 31P-31P COSY (GPQF) cosygpqf 2D 31P-31P COSY experiment using gradient pulses (magnitude mode) TBO5

Footnotes

#
Pulse sequence adapted for use with a spectrometer equipped with a QNP switch. The essence of the sequence is exactly the same as its non-QNP counterpart.
$
Due to the (adiabatic) broadband-decoupling during the acquisition time (AQ), the latter is restricted to a maximum of 0.4 sec for the Bruker Avance III HD 400 and the Bruker Avance Neo 600 and 0.25 sec for the Bruker Avance II+ 600 with a total decoupling power level of maximum 5 W. Note that the former also affects the maximum TD. A minimum D1 relaxation time of 5 sec is required.
+
Due to the broadband-decoupling during the acquisition time (AQ (F2)), the latter is restricted to a maximum of 0.2 sec. Note that this also affects the maximum TD (F2).

For an explanation of the two-character pulse program abbreviations used, please refer to the list of experiment abbreviations.