House rules

Access to the NMR Core

  • Unsupervised access to the NMR Core Facilities is restricted to trained staff and students, that obtained explicit permission to enter the NMR Core and hence, whose staff or student card is activated on the badge reader of the NMR Core.
  • For high-throughput NMR users unsupervised access is restricted to working hours (08:00-20:00 h from Monday to Friday) and to the high-throughput area. Access in the weekend is strictly forbidden.
  • As a high-throughput NMR user you are allowed to take along one staff member or student, for which you will personally perform measurements. You are obliged to supervise such accompanying person at all times and you will be held responsible for their activities.
  • The research NMR area is a prohibited area. Access to the research NMR area is only allowed upon explicit permission of a member of the NMR team (Gert Steurs (Department of Chemistry) or Dirk Dom (COK-CAT)). People allowed to enter the research area are listed on the access control sheet next to the door of this area.
  • Every activated user is obliged to badge upon access with their own personal badge and is expected to prevent unsupervised people from entering, when he/she enters or exits the NMR Core.
  • Using another person's badge or lending your badge to somebody else to enter the Core is strictly forbidden.
  • Problems or anything unusual - either with the NMR equipment or the rooms - should immediately be reported to a member of the NMR team.
  • Failure to adhere to these guidelines or instructions of a member of the NMR team will result in immediate and long-term termination of your access to the NMR Core.

Safety

Magnetic fields

A magnetic field surrounds the magnet in all directions. This field, known as the stray field, is invisible and does not get blocked by walls, floors or ceilings. Objects made of ferromagnetic materials, e.g. iron, steel etc. will be attracted to the magnet. Due to the active shielding of the superconducting coils, the stray magnetic field gradient is rather strong. Hence, caution must be taken to avoid bringing ferromagnetic objects close to the magnet. If a ferromagnetic object is brought too close, it may suddenly be drawn into the magnet with surprising force. This may damage the magnet, or cause personal injury to anybody in the way! Note that, in spite of the active shielding, the stray magnetic field directly above and directly below the magnet is very high and the attractive forces on ferromagnetic objects are very strong!

Due to the presence of these magnetic fields, persons with electronic, electrical or mechanical medical implants, such as cardiac pacemakers, biostimulators, and neurostimulators are not allowed in the NMR Core Facilities.

For the user's safety, a safety perimeter is marked on the floor with red tape around each magnet system. On top of that, the 5 Gauss (0.5 mT) stray field contour line is marked on the floor with black tape.

5 Gauss line
5 Gauss line represented by black tape.

Within the 5 Gauss radius (black markings), the following applies:

  • Small steel objects (keys, screwdrivers, bolts, paper clips, etc.) cannot be placed within this radius. These could cause serious damage if drawn into the magnet bore, especially when no probe is inserted in the magnet.
  • Mechanical watches may be damaged; digital watches can be worn safely.
  • Tape recorders, cameras and cell phones may be magnetized and irreparably damaged.
  • The stray field may erase information stored on magnetic tapes or discs. Bank cards, security passes or any devices containing a magnetic strip may be damaged.
  • CD/DVD’s will not be damaged (although CD/DVD drives may contain magnetic parts).
  • The color display of computer monitors may suffer some distortion when located too close to the magnet, although permanent damage is unlikely.
  • The operation of electronic, electrical or mechanical medical implants, such as cardiac pacemakers, biostimulators, and neurostimulators may be affected or even stopped in the presence of either static or changing magnetic fields. Not all pacemakers respond in the same way or at the same field strength if exposed to fields above 5 Gauss.
  • Besides electronic, electrical, and mechanical medical implants, other medical surgical implants such as aneurysm clips, surgical clips or prostheses, may contain ferromagnetic materials and therefore would be subject to strong attractive forces near to the NMR magnet system. This could result in injury or death. Additionally, in the vicinity of rapidly changing fields (e.g. pulsed gradient fields) eddy currents may be induced in the implant, hence resulting in heat generation and possibly create a life-threatening situation.
  • Electrical transformers in electronic devices may become magnetically saturated in fields above 5 mT (50 Gauss). The safety characteristics of equipment may also be affected.
Safety symbols
Safety symbols visible in the NMR Core.

Therefore, no person should ever step within or lean over these markings at any time!

Please note the following exceptions:

  • During maintenance or when a probe is being changed, trained personnel is allowed to cross the safety radius of the 5 Gauss line.
  • For the Bruker UltraShieldTM 600 magnet (part of the Bruker Avance II+ 600 setup), the 5 Gauss exclusion zone is too large, and crossing this line is required to place a sample in the sample changer. Even in this case, however, keep away as far as possible from the magnet (in a safe way) and take into account all the above mentioned points!

Cryogenic liquids

No touching symbol

In case of a magnet quench, the cryogenic liquids in the cryostat (LHe and LN2) will start to boil off vigorously and be released into the room via the emergency exhausts on the magnet system. The very large increase in volume accompanying the vaporization of the liquid into gas and the subsequent process of warming up is approximately 740:1 for helium and 680:1 for nitrogen. Although these gases are inert, if generated in large enough quantities, they can create dangerous circumstances if they displace the oxygen in the room.

Although the NMR Core Facilities are equipped for such emergencies, in case of a quench, everyone (both students and personnel) should leave the facilities as soon as possible in an orderly fashion. If necessary, inform one of the NMR Core Facilities staff members and notify Central Dispatch (2222).

Risk assessments

For use and maintenance of the high-field and low-field spectrometers, the necessary risk assessments (RAs) have been drafted and approved by the HSE Services. These include assessments regarding both chemical safety (CS) and working with machines, apparatuses and test setups (MATs). Users of the NMR Core facilities are obliged to examine these assessments and abide by them at all times. Furthermore, a reference to these RAs should be mentioned in the lab notebooks for experiments involving NMR. Note that only generic risks were recorded in the risk assessments, and additional information or risks should be evaluated in a seperate RA.

Please note that the PDF versions have not necessarily been updated to the latest version. If possible, download the latest version directly from the HSE Core Facilities share point website (login required).

Spectrometer status and mode

Each spectrometer is equipped with a 5-color tower light, placed next to the corresponding acquisition computer. Before logging on to the system, users must first check the status and running mode of the spectrometer. The upper three colors (red, yellow and green) indicate the status of the spectrometer, while the bottom two colors (white and blue) correspond to the operation mode. The color codes and their implications are listed below.

Spectrometer status
  • Spectrometer offline

    DO NOT USE te spectrometer or the acquisition computer. DO NOT SUBMIT samples. DO NOT take samples out of the sample changer.

  • Performance issues, external clients, filling cryogenics, ...

    You can use the spectrometer, but acquisition may be delayed or data may be sub-optimal.

  • Spectrometer online

    The spectrometer is fully operational. You can use the spectrometer and submit samples.

Operation mode
  • Automation mode

    Only people that have passed the training for AUTOMATION OPERATION (Icon-NMR Automation or Routine software) are allowed to use the spectrometer.

  • Manual mode

    Only people that have passed the training for MANUAL OPERATION (without automation software) are allowed to use the spectrometer.

Sample preparation

Compound

High-throughput NMR is suited for samples in the millimolar range and higher. For small organic compounds (Mw < 1000 Da), a regular 1H spectrum requires around 2-15 mg, while a 13C spectrum requires around 10-40 mg. Larger compounds (e.g. polymers or macromolecules) generally require considerably more material. The sensitivity of a nucleus is also inversely proportional to the required sample concentration. For a good indication of the sensitivity of all NMR active nuclei (represented by their receptivities relative to 1H and 13C), consult the interactive NMR isotopes table. Note that the resolution of your spectrum will increase with more diluted samples, whereas the sensitivity (or the signal-to-noise ratio) will increase with more concentrated samples.

Please note that these amounts are general guidelines, which should lead to efficient measurement of your sample using the standard acquisition parameters in automation. Take also into account that, if your sample is too diluted, it may be more efficient to measure at higher field, instead of increasing the number of scans and consuming a large amount of measuring time.

Samples containing corrosive, (possibly) explosive or (possibly) volatile compounds are not allowed to be inserted in the magnet bore without explicit permission from Gert Steurs. These samples could permanently corrode and destroy the probe or magnet. Examples of such samples include solutions with HF, HCl, azides, highly nitrated compounds, etc. For all samples containing volatile compounds, extreme care should be taken so that the pressure in the tube does not exceed the critical value for the tube. Tubes that break or explode inside the probe due to overpressure will permanently destroy the probe! On top of that, volatile products may also be released into the air through the cap of an NMR tube. Therefore, for all samples containing (possibly) volatile compounds (i.e. solvent or analyte) NMR tubes with a screw cap must be used and teflon tape must be applied between the cap and thread! These samples should be removed from the NMR Core as soon as possible after the measurement has finished.

Also samples containing paramagnetic species (Fe(0), Fe(III), Ni(II), ...) are not allowed to be inserted in the magnet bore without explicit permission from Gert Steurs. Such samples interfere with the magnetic field of the shim coils and may lead to permanent damage of the latter.

Solvent

The sample should be dissolved in deuterated solvent. Without deuterated solvent, the sample cannot be locked and shimmed, and therefore not measured in automation. The required sample height is 4.2 cm. For a 5 mm tube, this corresponds to a sample volume of 500-600 µL, depending on the wall thickness of the tube. Samples with less or more solvent will be badly shimmed, resulting in distortion of the peaks in your spectrum. For a good indication, download and print a 1:1 scaled image of a properly filled sample (print on 'Actual size').

Likewise, the presence of solid particles (or a biphasic liquid) in your sample is detrimental for the shimming process. Therefore, the optimal deuterated solvent should be chosen, ensuring complete dissolution of your compound and guaranteeing full homogeneity of the sample. Note that the shimming process is facilitated by diluted samples.

Take into account that bad shim settings will not only affect your own sample, but also the samples that are next in the queue. Please be aware that bad sample preparation affects other users’ measurements as well.

Preparation of the sample

As stated above, any form of inhomogeneity present in your sample, will result in improper shim settings. To ensure full homogeneity of your sample, the compound should be dissolved in the deuterated solvent in a secondary vial. If precipitation occurs, pass the sample through a syringe filter (with a pore size of 0.45 µm or smaller) before transferring it into the NMR tube.

NMR tubes

Use only NMR tubes that are fully intact. If a tube is cracked, broken, chipped, shortened or bent, throw it away immediately. In no case should such a tube be used anymore! Note that drying your NMR tubes in a high-temperature oven may bend the tubes. It is advised to dry your NMR tubes in a low-temperature (vacuum) oven.

KU Leuven members may consult our list of recommended NMR tubes. Note that for volatile and/or corrosive compounds or for measurements conducted at temperatures other than 298 K, specific tubes must be used.

Sample handling and submission

Every sample needs to be labeled clearly and properly at the upper side of the NMR tube with the initials of the head of the research group, your personal initials and a unique sample code (see section Experiment submission - Sample name). Make sure the ink of the written label stays well above the spinner and in no case comes out below the spinner! Do not put tape or paper labels, written by hand, on the NMR tubes to label your samples. Only labels made with a specific NMR label printer are allowed. When directly writing on the tube to label your sample, only write on the upper part of the tube, sticking out above the spinner.

Sample submission

To submit your sample into the automation queue, insert the sample in a spinner and adjust the height of the tube using the sample depth gauge. Note that a difference of 1 mm can mess up the shimming of your sample and those of the users after you. Make sure to clean the outside of the NMR tube and the spinner with a clean tissue before placing it in the sample changer or on top of the magnet bore (for manual measurements or measurements without a sample changer). This prohibits dirt from the outside of the tube (e.g. grease from your hands) to be deposited inside of the NMR probe. When grabbing the sample, either hold the spinner with a tissue or hold the upper part of the NMR tube sticking out of the spinner. In this way, both the spinner and the lower part of the tube (that is inserted in the probe) will stay clean.

In order to find an available holder in the sample changers, completely finished samples from other users may be removed. These samples should be placed in the corresponding plastic recipient on the white cupboards in the higthroughput area, containing all finished samples. Empty spinners should be placed back in the sample changer it was taken from and should never be removed or taken away. Please do not interchange spinners from the different sample changers.

Experiment submission

Sample name

When naming your sample (i.e. dataset) in Icon-NMR, keep the following in mind.

  • All sample names must refer to the person submitting the sample (which may not necessarily be the owner of the sample).
  • Always start the name of the sample with the initials of your promotor (head of the group whose user ID you use to login into Icon-NMR), followed by your full name initials (as provided on your access request form). Then continue with an appropriate and clear description of your sample. In conjunction with the previous point, this means that the beginning of each sample name should refer to the promotor and name of the person submitting the sample! The promotor initials, submittor initials and sample name should be separated by hyphens (-) or underscores (_).
  • Never use the same experiment number (expno) twice for the same sample name. In some cases Icon-NMR will warn you that the dataset already exists. This is, however, not always true. If the dataset name already existed, the new file may not be synchronized to the shared data drive or may overwrite a previous file. Note that there is no distinction between capital and small letters.
  • At risk of losing your data during the synchronization to the shared data drive, avoid using the following characters in the sample name: /, \, |, :, &, >, <, ', ", @. Preferably, names only contain letters and numbers and spaces are replaced by hyphens (-) or underscores (_). In any case, never end the name of your sample with a dot (.)!

For example: Felix Bloch, a PhD student in the group of prof. Edward Purcell, submits a sample for his colleague, Richard Ernst. Therefore, the sample name should start with EP-FB and may be something along the lines of EP-FB-RE-Sample1-CDCl3.

Number of samples

Maximum allowed number of sample holder slots that may be occupied by the same user at the same time (day + night + weekend experiments).

* If other users are waiting, limit your experiments to two consecutive samples, each taking no more than 15 min of total experiment time per sample.
Spectrometer Maximum number of slots
Avance 300 N/A*
Avance III HD 400 2 slots
Avance II+ 600 1 slot
Avance Neo 600 2 slots

Experiment time

All the available experiments in Icon-NMR are offered with a set of parameters that should perform well in 99% of the cases, provided that the samples are properly prepared and suited for NMR of course. Therefore, also the standard measuring time for each type of experiment should suffice. However, in some cases, users may want to modify some of the standard parameters. The parameters that can be changed are listed below. In case of doubt, contact Gert Steurs.

* Only availble upon specific request and for single experiments.
Parameter Meaning Bruker Avance 300 Bruker Avance III HD 400 Bruker Avance II+ 600 Bruker Avance Neo 600
1TD Size of the FID in the F1 dimension
2TD Size of the FID in the F2 dimension
AQ Acquisition time (sec)
DS Number of dummy scans
NS Number of scans
1SW Spectral width in the F1 dimension (ppm)
2SW Spectral width in the F2 dimension (ppm)
O1P Middle of the spectrum for NUC1 (ppm)
D1 Relaxation delay (sec, should be 1-5 * T1)
D6 Delay for evolution of long-range coupling (sec, used in LR-COSY, should be 1/2J)
D8 Mixing time (sec, used in NOESY experiments)
FNTYPE nD acquisition mode (0 for traditional; 2 for NUS)
NUSAMOUNT Amount of sparse sampling (%)
TE Sample temperature (K, 283-353 K) * * *
usera5 User acquisition parameter 5
Remarks
  • Increasing the number of scans only increases the signal-to-noise ratio (= sensitivity), not the resolution.
  • Mass-limited samples that do not show a decent 1H spectrum within 30 minutes probably need another technique to be analyzed or have to be recorded at higher field. In this case, it also does not make sense to start recording the 13C spectrum of such dilute samples.

Be aware that modifying the experiment parameters may have an influence on the measuring time. Therefore, for each type of experiment, a maximum allowed measuring time has been determined. Note that when the maximum time is exceeded, the submitted experiment will automatically be aborted by the software during an initial check at runtime of the experiment without any further warning (hence, the experiments will appear in the Icon-NMR queue and will be cancelled only at the moment the sample is about to be measured). The maximum experiment time (in minutes) for each experiment is listed in the table below.

#
Pulse sequence adapted for use with a spectrometer equipped with a QNP switch. The essence of the sequence is exactly the same as its non-QNP counterpart.
$
Due to the (adiabatic) broadband-decoupling during the acquisition time (AQ), the latter is restricted to a maximum of 0.4 sec for the Bruker Avance III HD 400 and the Bruker Avance Neo 600 and 0.25 sec for the Bruker Avance II+ 600 with a total decoupling power level of maximum 5 W. Note that the former also affects the maximum TD. A minimum D1 relaxation time of 5 sec is required.
+
Due to the broadband-decoupling during the acquisition time (AQ (F2)), the latter is restricted to a maximum of 0.2 sec. Note that this also affects the maximum TD (F2).
Experiment AV 300 AV III HD 400 AV II+ 600 AV Neo 600
Day Night Weekend Day Night Weekend Day Night Weekend Day Night Weekend
1D 1H (30) 5 10 30 5 10 30 5 10 30 5 10 30
1D 1H 5 10 30 5 10 30 5 10 30 5 10 30
1D 1H QNMR (30) - - - 5 10 30 - - - - - -
1D 1H (PR) 5 10 30 5 10 30 5 10 30 5 10 30
1D 1H (GPPR1D) - - - 5 10 30 5 10 30 5 10 30
1D 1H (ESGP) - - - 5 10 30 5 10 30 5 10 30
1D 1H{13C} (IG30ADIA)$ - - - 5 10 30 5 10 30 5 10 30
1D 1H{19F} (IG30ADIA) (QNP)#, $ - - - 5 10 30 - - - - - -
1D 1H{19F} (IG30ADIA)$ - - - - - - - - - 5 10 30
1D 1H{31P} (IG30ADIA)$ - - - 5 10 30 5 10 30 5 10 30
1D 1H{13C,19F} (IG30ADIA)$ - - - - - - - - - 5 10 30
1D 2H (LOCK) - - - 5 10 30 5 10 30 5 10 30
1D 7Li (30) - - - 5 10 30 5 10 30 5 10 30
1D 11B - - - 5 10 30 5 10 30 5 10 30
1D 11B{1H} (PG) - - - 5 10 30 5 10 30 5 10 30
1D 13C (GD30) 5 60 180 5 60 180 5 60 180 5 60 180
1D 13C{1H} (PG30) 5 60 180 5 60 180 5 60 180 5 60 180
1D 13C{1H} (IG30) 5 60 180 5 60 180 5 60 180 5 60 180
1D 13C{1H} DEPT135 (SP) 5 30 90 5 30 90 5 30 90 5 30 90
1D 13C{1H} DEPT90 (SP) 5 30 90 5 30 90 5 30 90 5 30 90
1D 13C{1H} DEPT45 (SP) 5 30 90 5 30 90 5 30 90 5 30 90
1D 13C{1H} DEPTQ135 (GPSP) - - - 5 30 90 5 30 90 5 30 90
1D 13C{1H} APT 5 30 90 5 30 90 - - - - - -
1D 15N - - - 5 60 180 5 60 180 5 60 180
1D 15N{1H} (IG) - - - 5 60 180 5 60 180 5 60 180
1D 19F (QNP)# - - - 5 10 30 - - - - - -
1D 19F - - - - - - - - - 5 10 30
1D 19F{1H} (IG) (QNP)# - - - 5 10 30 - - - - - -
1D 19F{1H} (IG) - - - - - - - - - 5 10 30
1D 27Al - - - 5 10 30 5 10 30 5 10 30
1D 29Si - - - 5 15 45 5 15 45 5 15 45
1D 29Si{1H} (IG) - - - 5 15 45 5 15 45 5 15 45
1D 31P (30) - - - 5 10 30 5 10 30 5 10 30
1D 31P{1H} (PG30) - - - 5 10 30 5 10 30 5 10 30
1D 31P{1H} (IG30) - - - 5 10 30 5 10 30 5 10 30
1D 35Cl - - - 5 30 90 5 30 90 5 30 90
1D 43Ca - - - - - - 5 60 180 5 60 180
1D 43Ca (ARING) - - - - - - 5 60 180 5 60 180
1D 51V - - - 5 10 30 5 10 30 5 10 30
1D 119Sn (30) - - - 5 10 30 5 10 30 5 10 30
1D 119Sn{1H} (IG30) - - - 5 10 30 5 10 30 5 10 30
1D 195Pt - - - 5 30 90 5 30 90 5 30 90
1D 195Pt{1H} - - - 5 30 90 5 30 90 5 30 90
2D 1H T1 (IR) - - - 5 30 90 5 30 90 5 30 90
2D 1H-1H COSY (GPPPQF) - - - 5 30 90 5 30 90 5 30 90
2D 1H-1H DQF-COSY (GPDFPHPP) - - - 5 30 90 5 30 90 5 30 90
2D 1H-1H LR-COSY (GPLRPPQF) - - - 5 30 90 5 30 90 5 30 90
2D 1H-1H NOESY (GPPHPP) - - - 5 60 180 5 60 180 5 60 180
2D 1H-1H ROESY (PHPP.2) - - - 5 60 180 5 60 180 5 60 180
2D 1H-1H DOSY (LEDBPGP2S) - - - 5 60 180 5 60 180 5 60 180
2D 1H-13C HSQC-DEPT (EDETGPSISP2.2ADIA)+ - - - 5 60 120 5 60 120 5 60 120
2D 1H-13C HSQC (ETGPSP.2ADIA)+ - - - 5 60 120 5 60 120 5 60 120
2D 1H-13C HSQC (ETGPSP.3)+ - - - 5 60 120 - - - - - -
2D 1H-13C HMBC (GPLPNDQF) - - - 5 60 180 5 60 180 5 60 180
2D 1H-13C HMBC (ETGPL3ND) - - - 5 60 180 5 60 180 5 60 180
2D 1H-15N HSQC (ETGPSI2)+ - - - 5 60 120 5 60 120 5 60 120
2D 1H-15N HMBC (GPNDQF) - - - 5 60 180 5 60 180 5 60 180
2D 1H-31P HMQC (ETGPSI.2ADIA)+ - - - 5 60 120 5 60 120 5 60 120
2D 1H-31P HMBC (GPLPNDQF) - - - 5 60 180 5 60 180 5 60 180
2D 1H-195Pt HMQC (QF)+ - - - 5 60 120 5 60 120 5 60 120
2D 19F-13C HMQC (ETGPSI.2)+ - - - - - - - - - 5 60 120
2D 19F-13C HMBC (GPLPNDQF) - - - - - - - - - 5 60 180
2D 19F-19F COSY (GPPPQF) - - - 5 30 90 - - - 5 30 90
2D 19F-19F NOESY (GPPHPP) - - - 5 60 180 - - - 5 60 180
2D 31P-31P COSY (GPQF) - - - 5 60 120 5 60 120 5 60 120

Day, night and weekend queue

Day experiments are run on week days (Mon-Fri) from 8:00 to 20:00 h, night experiments on week days from 20:00 to 8:00 h and during the weekend (Fri 20:00 h – Mon 8:00 h). Experiments with a measuring time of 5 minutes or less may be submitted as a day experiment (sun icon). When the time exceeds 5 minutes, the measurement must be submitted as a night experiment (moon icon). Weekend measurements are a special type of night experiments (using also the moon icon), which are run from Friday 20:00 h until Monday 08:00 h. These experiments may only be submitted on Friday after 17:00 h.

Avance 300 (no sample changer)

As the spectrometer is running solely in Routine mode, only one sample can be submitted at the same time. For submission of a night or weekend experiment on this spectrometer, please contact the NMR Core administrator team.

Avance III HD 400, Avance Neo 600 (sample changer with 60 positions)

During the week, sample holder positions 2 to 24 are reserved for night experiments (or combined day + night experiments). Positions 25 to 60 can only be used for day experiments only.

From Friday 17:00 h, all sample holder positions (2 to 60) may be used for night (= weekend) experiments. However, the total time of the night queue may not exceed 55 hours. All other limitations (e.g. time limit per experiment, total number of samples, etc.) still apply as well.

Sample position 1 is always reserved for Autocalibrate. This position cannot be used and the sample must never be removed.

Avance II+ 600 (sample changer with 24 positions)

During the week, sample holder positions 2 to 12 are reserved for night experiments (or combined day + night experiments). Positions 13 to 24 can only be used for day experiments only.

From Friday 17:00 h, all sample holder positions (2 to 24) may be used for night (= weekend) experiments. However, the total time of the night queue may not exceed 55 hours. All other limitations (e.g. time limit per experiment, total number of samples, etc.) still apply as well.

Sample position 1 is always reserved for Autocalibrate. This position cannot be used and the sample must never be removed.

Total queue time

Maximum allowed total measuring time of all submitted samples together per user in the queue.

Day queue Week night queue Weekend queue
30 minutes 2 hours 6 hours

Measurements that take longer need to be motivated and can only be performed during the weekend after approval via email to priority[at]leuvennmrcore[dot]be.

Priority measurements

Priority measurements and measurements that use the 'set start time' option always need to be approved on beforehand via email to priority[at]leuvennmrcore[dot]be. Indicate your number of samples, type(s) of measurement(s), estimated required time and when you would like to perform the experiment(s).

Exceptions

Deviations from these rules are possible, but only after approval via email to priority[at]leuvennmrcore[dot]be.

Camera surveillance

Since october 2020, the Leuven NMR Core was permitted by KU Leuven to equip the HT area with camera surveillance. The camera is used for remote monitoring and maintenance of the equipment and will only be used as such. The device is operational 24/7, but the live-feed is only monitored for a select time period, during which the remote maintenance or intervention is performed. Only a limited number of Leuven NMR Core administrators has access to this feed. Images are neither recorded nor stored.

Acknowledging the Leuven NMR Core

All users are obliged to acknowledge the the proper instances when reporting or publishing data recorded in the NMR Core facilities. More information about this subject can be found in the Author Guidelines.

Policy on Data Usage and Commercial Service Provision

Leuven NMR Core provides state-of-the-art instrumentation to support the academic and research missions of KU Leuven. Our internal rates are intended exclusively for non-profit research conducted by KU Leuven faculty, staff, and students.

The use of our facilities for the purpose of generating commercial data, providing service measurements for third parties, or performing contract research for external entities (whether in exchange for monetary payment, goods, services, or other considerations) is strictly prohibited.

All data acquired in this facility must be used solely for internal research projects only. Any external organization, commercial company, or third party interested in utilizing our NMR services must contract directly with the facility via info[at]leuvennmrcore[dot]be. Any user found to be brokering NMR time or data for unauthorized commercial gain may face suspension of facility access and may be subject to further disciplinary action.

By submitting samples to the Leuven NMR Core facilities, you agree to abide by these terms. If you have questions regarding whether your project qualifies for internal pricing, please contact info[at]leuvennmrcore[dot]be.

Incompliance

Any violation of the rules listed above may result in removal of all your samples from the queue. Repeated incompliance with the NMR House Rules will result in immediate (temporary) termination of your access to the NMR Core and/or further disciplinary action.